This page is part of the CardX - Cardiac Implantable Electronic Devices (v1.0.0-ballot: STU 1 Ballot 1) based on FHIR (HL7® FHIR® Standard) v5.0.0. No current official version has been published yet. For a full list of available versions, see the Directory of published versions
Official URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-connectivity-status | Version: 1.0.0-ballot | |||
Standards status: Trial-use | Maturity Level: 2 | Computable Name: CIEDConnectivityStatus |
Profile of the observation resource to present the most recent connectivity status and associated information for a CIED/Monitor pair.
CIEDs and monitors connect at regular intervals to exchange data ranging from every few minutes to every other week depending on the device/monitor configuration and manufacturer. The connectivity observation profile is a snapshot in time and is not intended to be used to query historical log information that may have been captured by the remote monitoring IoT platform. A single observation will be returned for every CIED/monitor pair.
Usages:
You can also check for usages in the FHIR IG Statistics
Description of Profiles, Differentials, Snapshots and how the different presentations work.
Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
---|---|---|---|---|
![]() ![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7, obs-8 |
![]() ![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created |
![]() ![]() ![]() |
0..* | Resource | Contained, inline Resources | |
![]() ![]() ![]() |
1..* | Extension | Extension Slice: Unordered, Open by value:url | |
![]() ![]() ![]() ![]() |
S | 1..* | Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) | Cardiac implantable device(s) URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension |
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored |
![]() ![]() ![]() |
?!SΣ | 1..1 | code | registered | preliminary | final | amended + Binding: ObservationStatus (required): Codes providing the status of an observation. |
![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: observation-cied-connectivity | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() |
SΣ | 1..1 | Reference(Patient - CIED Patient Profile) | Who and/or what the observation is about |
![]() ![]() ![]() |
SΣ | 1..1 | dateTime | Clinically relevant time/time-period for observation |
![]() ![]() ![]() |
ΣC | 0..1 | Actual result | |
![]() ![]() ![]() ![]() |
Quantity | |||
![]() ![]() ![]() ![]() |
CodeableConcept | |||
![]() ![]() ![]() ![]() |
string | |||
![]() ![]() ![]() ![]() |
boolean | |||
![]() ![]() ![]() ![]() |
integer | |||
![]() ![]() ![]() ![]() |
Range | |||
![]() ![]() ![]() ![]() |
Ratio | |||
![]() ![]() ![]() ![]() |
SampledData | |||
![]() ![]() ![]() ![]() |
time | |||
![]() ![]() ![]() ![]() |
dateTime | |||
![]() ![]() ![]() ![]() |
Period | |||
![]() ![]() ![]() ![]() |
Attachment | |||
![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||
![]() ![]() ![]() |
S | 0..1 | Reference(Device - CIED Monitor Profile) | A reference to the device that generates the measurements or the device settings for the device |
![]() ![]() ![]() |
SΣC | 1..* | BackboneElement | CIED Connectivity Components Slice: Unordered, Open by value:code |
![]() ![]() ![]() ![]() |
Content/Rules for all slices | |||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | Actual component result | |
![]() ![]() ![]() ![]() ![]() ![]() |
Quantity | |||
![]() ![]() ![]() ![]() ![]() ![]() |
CodeableConcept | |||
![]() ![]() ![]() ![]() ![]() ![]() |
string | |||
![]() ![]() ![]() ![]() ![]() ![]() |
boolean | |||
![]() ![]() ![]() ![]() ![]() ![]() |
integer | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Range | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Ratio | |||
![]() ![]() ![]() ![]() ![]() ![]() |
SampledData | |||
![]() ![]() ![]() ![]() ![]() ![]() |
time | |||
![]() ![]() ![]() ![]() ![]() ![]() |
dateTime | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Period | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Attachment | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||
![]() ![]() ![]() ![]() |
SΣC | 1..1 | BackboneElement | The overall system connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-status | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Value Set (required) |
![]() ![]() ![]() ![]() |
SΣC | 0..* | BackboneElement | Additional modifiers to connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-modifier | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Modifier Value Set (required) |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing reasons why the patient may be listed at a given status and/or modifier. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-reason | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing steps to get the patient to a connected status |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-guidance | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-cied-connectivity-time | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-monitor-connectivity-time | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-cied-connectivity-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-monitor-connectivity-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the most recent remote device interrogation with a personal monitor. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-interrogation-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-interrogation-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-secondary-interrogation-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time interval in days between expected CIED and remote monitoring software communication. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connection-interval | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time period in days in which successful communication must take place for the connectivity status to remain connected. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: expected-time-period | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result |
![]() |
Path | Conformance | ValueSet / Code | URI |
Observation.status | required | ObservationStatushttp://hl7.org/fhir/ValueSet/observation-status|5.0.0 From the FHIR Standard | |
Observation.code | example | Pattern Value: observation-cied-connectivityhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component.code | example | LOINCCodes(a valid code from LOINC)http://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityStatus.code | example | Pattern Value: connectivity-statushttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityStatus.value[x] | required | CIEDConnectivityStateVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-vs From this IG | |
Observation.component:connectivityModifier.code | example | Pattern Value: connectivity-modifierhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityModifier.value[x] | required | CIEDConnectivityStateModifierVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-modifier-vs From this IG | |
Observation.component:statusReason.code | example | Pattern Value: status-reasonhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:statusGuidance.code | example | Pattern Value: status-guidancehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.code | example | Pattern Value: last-cied-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.code | example | Pattern Value: last-monitor-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.code | example | Pattern Value: next-cied-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.code | example | Pattern Value: next-monitor-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.code | example | Pattern Value: last-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.code | example | Pattern Value: next-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.code | example | Pattern Value: next-secondary-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectionInterval.code | example | Pattern Value: connection-intervalhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:expectedTimePeriod.code | example | Pattern Value: expected-time-periodhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard |
Id | Grade | Path(s) | Details | Requirements |
dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources : contained.contained.empty() | |
dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource : contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty() | |
dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated : contained.meta.versionId.empty() and contained.meta.lastUpdated.empty() | |
dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label : contained.meta.security.empty() | |
dom-6 | best practice | Observation | A resource should have narrative for robust management : text.`div`.exists() | |
ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children : hasValue() or (children().count() > id.count()) | |
ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both : extension.exists() != value.exists() | |
obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present : dataAbsentReason.empty() or value.empty() | |
obs-7 | error | Observation | If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). : value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty() | |
obs-8 | error | Observation | bodyStructure SHALL only be present if Observation.bodySite is not present : bodySite.exists() implies bodyStructure.empty() |
This structure is derived from Observation
Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
---|---|---|---|---|
![]() ![]() |
0..* | Observation | Measurements and simple assertions | |
![]() ![]() ![]() |
1..* | Extension | Extension Slice: Unordered, Open by value:url | |
![]() ![]() ![]() ![]() |
S | 1..* | Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) | Cardiac implantable device(s) URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension |
![]() ![]() ![]() |
S | 1..1 | code | registered | preliminary | final | amended + |
![]() ![]() ![]() |
S | 1..1 | CodeableConcept | Type of observation (code / type) Required Pattern: At least the following |
![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: observation-cied-connectivity | |
![]() ![]() ![]() |
S | 1..1 | Reference(Patient - CIED Patient Profile) | Who and/or what the observation is about |
![]() ![]() ![]() |
S | 1..1 | dateTime | Clinically relevant time/time-period for observation |
![]() ![]() ![]() |
S | 0..1 | Reference(Device - CIED Monitor Profile) | A reference to the device that generates the measurements or the device settings for the device |
![]() ![]() ![]() |
S | 1..* | BackboneElement | CIED Connectivity Components Slice: Unordered, Open by value:code |
![]() ![]() ![]() ![]() |
Content/Rules for all slices | |||
![]() ![]() ![]() ![]() ![]() |
S | 1..1 | CodeableConcept | Type of component observation (code / type) |
![]() ![]() ![]() ![]() ![]() |
S | 0..1 | Quantity, CodeableConcept, string, boolean, integer, Range, Ratio, SampledData, time, dateTime, Period, Attachment, Reference(MolecularSequence) | Actual component result |
![]() ![]() ![]() ![]() |
S | 1..1 | BackboneElement | The overall system connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-status | |
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Value Set (required) | |
![]() ![]() ![]() ![]() |
S | 0..* | BackboneElement | Additional modifiers to connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-modifier | |
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Modifier Value Set (required) | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | Text describing reasons why the patient may be listed at a given status and/or modifier. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-reason | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | Text describing steps to get the patient to a connected status |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-guidance | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the most recent communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-cied-connectivity-time | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the most recent communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-monitor-connectivity-time | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the next expected communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-cied-connectivity-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the next expected communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-monitor-connectivity-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The date of the most recent remote device interrogation with a personal monitor. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-interrogation-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-interrogation-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-secondary-interrogation-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The time interval in days between expected CIED and remote monitoring software communication. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connection-interval | |
![]() ![]() ![]() ![]() ![]() |
0..1 | integer | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The time period in days in which successful communication must take place for the connectivity status to remain connected. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: expected-time-period | |
![]() ![]() ![]() ![]() ![]() |
0..1 | integer | Actual component result | |
![]() |
Path | Conformance | ValueSet | URI |
Observation.component:connectivityStatus.value[x] | required | CIEDConnectivityStateVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-vs From this IG | |
Observation.component:connectivityModifier.value[x] | required | CIEDConnectivityStateModifierVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-modifier-vs From this IG |
Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() | ||||
---|---|---|---|---|---|---|---|---|
![]() ![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7, obs-8 | ||||
![]() ![]() ![]() |
Σ | 0..1 | id | Logical id of this artifact | ||||
![]() ![]() ![]() |
Σ | 0..1 | Meta | Metadata about the resource | ||||
![]() ![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created | ||||
![]() ![]() ![]() |
0..1 | code | Language of the resource content Binding: AllLanguages (required): IETF language tag for a human language
| |||||
![]() ![]() ![]() |
0..1 | Narrative | Text summary of the resource, for human interpretation This profile does not constrain the narrative in regard to content, language, or traceability to data elements | |||||
![]() ![]() ![]() |
0..* | Resource | Contained, inline Resources | |||||
![]() ![]() ![]() |
1..* | Extension | Extension Slice: Unordered, Open by value:url | |||||
![]() ![]() ![]() ![]() |
S | 1..* | Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) | Cardiac implantable device(s) URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension | ||||
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored | ||||
![]() ![]() ![]() |
Σ | 0..* | Identifier | Business Identifier for observation | ||||
![]() ![]() ![]() |
Σ | 0..1 | canonical(ObservationDefinition), Reference(ObservationDefinition) | Instantiates FHIR ObservationDefinition | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(CarePlan | DeviceRequest | ImmunizationRecommendation | MedicationRequest | NutritionOrder | ServiceRequest) | Fulfills plan, proposal or order | ||||
![]() ![]() ![]() |
0..* | BackboneElement | Triggering observation(s) | |||||
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() |
Σ | 1..1 | Reference(Observation) | Triggering observation | ||||
![]() ![]() ![]() ![]() |
Σ | 1..1 | code | reflex | repeat | re-run Binding: TriggeredBytype (required): The type of TriggeredBy Observation. | ||||
![]() ![]() ![]() ![]() |
0..1 | string | Reason that the observation was triggered | |||||
![]() ![]() ![]() |
Σ | 0..* | Reference(MedicationAdministration | MedicationDispense | MedicationStatement | Procedure | Immunization | ImagingStudy | GenomicStudy) | Part of referenced event | ||||
![]() ![]() ![]() |
?!SΣ | 1..1 | code | registered | preliminary | final | amended + Binding: ObservationStatus (required): Codes providing the status of an observation. | ||||
![]() ![]() ![]() |
0..* | CodeableConcept | Classification of type of observation Binding: ObservationCategoryCodes (preferred): Codes for high level observation categories. | |||||
![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: observation-cied-connectivity | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() |
SΣ | 1..1 | Reference(Patient - CIED Patient Profile) | Who and/or what the observation is about | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(Resource) | What the observation is about, when it is not about the subject of record | ||||
![]() ![]() ![]() |
Σ | 0..1 | Reference(Encounter) | Healthcare event during which this observation is made | ||||
![]() ![]() ![]() |
SΣ | 1..1 | dateTime | Clinically relevant time/time-period for observation | ||||
![]() ![]() ![]() |
Σ | 0..1 | instant | Date/Time this version was made available | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(Practitioner | PractitionerRole | Organization | CareTeam | Patient | RelatedPerson) | Who is responsible for the observation | ||||
![]() ![]() ![]() |
ΣC | 0..1 | Actual result | |||||
![]() ![]() ![]() ![]() |
Quantity | |||||||
![]() ![]() ![]() ![]() |
CodeableConcept | |||||||
![]() ![]() ![]() ![]() |
string | |||||||
![]() ![]() ![]() ![]() |
boolean | |||||||
![]() ![]() ![]() ![]() |
integer | |||||||
![]() ![]() ![]() ![]() |
Range | |||||||
![]() ![]() ![]() ![]() |
Ratio | |||||||
![]() ![]() ![]() ![]() |
SampledData | |||||||
![]() ![]() ![]() ![]() |
time | |||||||
![]() ![]() ![]() ![]() |
dateTime | |||||||
![]() ![]() ![]() ![]() |
Period | |||||||
![]() ![]() ![]() ![]() |
Attachment | |||||||
![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||||||
![]() ![]() ![]() |
C | 0..1 | CodeableConcept | Why the result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | ||||
![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() |
0..* | Annotation | Comments about the observation | |||||
![]() ![]() ![]() |
C | 0..1 | CodeableConcept | Observed body part Binding: SNOMEDCTBodyStructures (example): SNOMED CT Body site concepts | ||||
![]() ![]() ![]() |
C | 0..1 | Reference(BodyStructure) | Observed body structure | ||||
![]() ![]() ![]() |
0..1 | CodeableConcept | How it was done Binding: ObservationMethods (example): Methods for simple observations. | |||||
![]() ![]() ![]() |
C | 0..1 | Reference(Specimen | Group) | Specimen used for this observation Constraints: obs-9 | ||||
![]() ![]() ![]() |
S | 0..1 | Reference(Device - CIED Monitor Profile) | A reference to the device that generates the measurements or the device settings for the device | ||||
![]() ![]() ![]() |
C | 0..* | BackboneElement | Provides guide for interpretation Constraints: obs-3 | ||||
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() |
C | 0..1 | SimpleQuantity | Low Range, if relevant | ||||
![]() ![]() ![]() ![]() |
C | 0..1 | SimpleQuantity | High Range, if relevant | ||||
![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Normal value, if relevant Binding: ObservationReferenceRangeNormalValueCodes (extensible): Codes identifying the normal value of the observation. | |||||
![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Reference range qualifier Binding: ObservationReferenceRangeMeaningCodes (preferred): Code for the meaning of a reference range. | |||||
![]() ![]() ![]() ![]() |
0..* | CodeableConcept | Reference range population Binding: ObservationReferenceRangeAppliesToCodes (example): Codes identifying the population the reference range applies to. | |||||
![]() ![]() ![]() ![]() |
0..1 | Range | Applicable age range, if relevant | |||||
![]() ![]() ![]() ![]() |
C | 0..1 | markdown | Text based reference range in an observation | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(Observation | QuestionnaireResponse | MolecularSequence) | Related resource that belongs to the Observation group | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(DocumentReference | ImagingStudy | ImagingSelection | QuestionnaireResponse | Observation | MolecularSequence | GenomicStudy) | Related resource from which the observation is made | ||||
![]() ![]() ![]() |
SΣC | 1..* | BackboneElement | CIED Connectivity Components Slice: Unordered, Open by value:code | ||||
![]() ![]() ![]() ![]() |
Content/Rules for all slices | |||||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. | ||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | Actual component result | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
Quantity | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
CodeableConcept | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
string | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
boolean | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
integer | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Range | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Ratio | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
SampledData | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
time | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
dateTime | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Period | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Attachment | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 1..1 | BackboneElement | The overall system connectivity status as calculated by the CIED manufacturer. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-status | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Value Set (required) | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..* | BackboneElement | Additional modifiers to connectivity status as calculated by the CIED manufacturer. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-modifier | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Modifier Value Set (required) | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing reasons why the patient may be listed at a given status and/or modifier. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-reason | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing steps to get the patient to a connected status | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-guidance | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication between the CIED and monitor. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-cied-connectivity-time | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication from the monitor to the remote monitoring platform. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-monitor-connectivity-time | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication between the CIED and monitor. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-cied-connectivity-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication from the monitor to the remote monitoring platform. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-monitor-connectivity-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the most recent remote device interrogation with a personal monitor. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-interrogation-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-interrogation-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-secondary-interrogation-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time interval in days between expected CIED and remote monitoring software communication. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connection-interval | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time period in days in which successful communication must take place for the connectivity status to remain connected. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: expected-time-period | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() |
Path | Conformance | ValueSet / Code | URI |
Observation.language | required | AllLanguageshttp://hl7.org/fhir/ValueSet/all-languages|5.0.0 From the FHIR Standard | |
Observation.triggeredBy.type | required | TriggeredBytypehttp://hl7.org/fhir/ValueSet/observation-triggeredbytype|5.0.0 From the FHIR Standard | |
Observation.status | required | ObservationStatushttp://hl7.org/fhir/ValueSet/observation-status|5.0.0 From the FHIR Standard | |
Observation.category | preferred | ObservationCategoryCodeshttp://hl7.org/fhir/ValueSet/observation-category From the FHIR Standard | |
Observation.code | example | Pattern Value: observation-cied-connectivityhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.bodySite | example | SNOMEDCTBodyStructureshttp://hl7.org/fhir/ValueSet/body-site From the FHIR Standard | |
Observation.method | example | ObservationMethodshttp://hl7.org/fhir/ValueSet/observation-methods From the FHIR Standard | |
Observation.referenceRange.normalValue | extensible | ObservationReferenceRangeNormalValueCodeshttp://hl7.org/fhir/ValueSet/observation-referencerange-normalvalue From the FHIR Standard | |
Observation.referenceRange.type | preferred | ObservationReferenceRangeMeaningCodeshttp://hl7.org/fhir/ValueSet/referencerange-meaning From the FHIR Standard | |
Observation.referenceRange.appliesTo | example | ObservationReferenceRangeAppliesToCodeshttp://hl7.org/fhir/ValueSet/referencerange-appliesto From the FHIR Standard | |
Observation.component.code | example | LOINCCodes(a valid code from LOINC)http://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:connectivityStatus.code | example | Pattern Value: connectivity-statushttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityStatus.value[x] | required | CIEDConnectivityStateVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-vs From this IG | |
Observation.component:connectivityStatus.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:connectivityStatus.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:connectivityModifier.code | example | Pattern Value: connectivity-modifierhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityModifier.value[x] | required | CIEDConnectivityStateModifierVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-modifier-vs From this IG | |
Observation.component:connectivityModifier.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:connectivityModifier.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:statusReason.code | example | Pattern Value: status-reasonhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:statusReason.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:statusReason.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:statusGuidance.code | example | Pattern Value: status-guidancehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:statusGuidance.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:statusGuidance.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.code | example | Pattern Value: last-cied-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.code | example | Pattern Value: last-monitor-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.code | example | Pattern Value: next-cied-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.code | example | Pattern Value: next-monitor-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.code | example | Pattern Value: last-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.code | example | Pattern Value: next-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.code | example | Pattern Value: next-secondary-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:connectionInterval.code | example | Pattern Value: connection-intervalhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectionInterval.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:connectionInterval.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:expectedTimePeriod.code | example | Pattern Value: expected-time-periodhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:expectedTimePeriod.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:expectedTimePeriod.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard |
Id | Grade | Path(s) | Details | Requirements |
dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources : contained.contained.empty() | |
dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource : contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty() | |
dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated : contained.meta.versionId.empty() and contained.meta.lastUpdated.empty() | |
dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label : contained.meta.security.empty() | |
dom-6 | best practice | Observation | A resource should have narrative for robust management : text.`div`.exists() | |
ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children : hasValue() or (children().count() > id.count()) | |
ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both : extension.exists() != value.exists() | |
obs-3 | error | Observation.referenceRange | Must have at least a low or a high or text : low.exists() or high.exists() or text.exists() | |
obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present : dataAbsentReason.empty() or value.empty() | |
obs-7 | error | Observation | If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). : value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty() | |
obs-8 | error | Observation | bodyStructure SHALL only be present if Observation.bodySite is not present : bodySite.exists() implies bodyStructure.empty() | |
obs-9 | error | Observation.specimen | If Observation.specimen is a reference to Group, the group can only have specimens : (reference.resolve().exists() and reference.resolve() is Group) implies reference.resolve().member.entity.resolve().all($this is Specimen) |
This structure is derived from Observation
Summary
Mandatory: 6 elements
Must-Support: 22 elements
Structures
This structure refers to these other structures:
Extensions
This structure refers to these extensions:
Slices
This structure defines the following Slices:
Maturity: 2
Key Elements View
Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
---|---|---|---|---|
![]() ![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7, obs-8 |
![]() ![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created |
![]() ![]() ![]() |
0..* | Resource | Contained, inline Resources | |
![]() ![]() ![]() |
1..* | Extension | Extension Slice: Unordered, Open by value:url | |
![]() ![]() ![]() ![]() |
S | 1..* | Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) | Cardiac implantable device(s) URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension |
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored |
![]() ![]() ![]() |
?!SΣ | 1..1 | code | registered | preliminary | final | amended + Binding: ObservationStatus (required): Codes providing the status of an observation. |
![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: observation-cied-connectivity | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() |
SΣ | 1..1 | Reference(Patient - CIED Patient Profile) | Who and/or what the observation is about |
![]() ![]() ![]() |
SΣ | 1..1 | dateTime | Clinically relevant time/time-period for observation |
![]() ![]() ![]() |
ΣC | 0..1 | Actual result | |
![]() ![]() ![]() ![]() |
Quantity | |||
![]() ![]() ![]() ![]() |
CodeableConcept | |||
![]() ![]() ![]() ![]() |
string | |||
![]() ![]() ![]() ![]() |
boolean | |||
![]() ![]() ![]() ![]() |
integer | |||
![]() ![]() ![]() ![]() |
Range | |||
![]() ![]() ![]() ![]() |
Ratio | |||
![]() ![]() ![]() ![]() |
SampledData | |||
![]() ![]() ![]() ![]() |
time | |||
![]() ![]() ![]() ![]() |
dateTime | |||
![]() ![]() ![]() ![]() |
Period | |||
![]() ![]() ![]() ![]() |
Attachment | |||
![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||
![]() ![]() ![]() |
S | 0..1 | Reference(Device - CIED Monitor Profile) | A reference to the device that generates the measurements or the device settings for the device |
![]() ![]() ![]() |
SΣC | 1..* | BackboneElement | CIED Connectivity Components Slice: Unordered, Open by value:code |
![]() ![]() ![]() ![]() |
Content/Rules for all slices | |||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | Actual component result | |
![]() ![]() ![]() ![]() ![]() ![]() |
Quantity | |||
![]() ![]() ![]() ![]() ![]() ![]() |
CodeableConcept | |||
![]() ![]() ![]() ![]() ![]() ![]() |
string | |||
![]() ![]() ![]() ![]() ![]() ![]() |
boolean | |||
![]() ![]() ![]() ![]() ![]() ![]() |
integer | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Range | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Ratio | |||
![]() ![]() ![]() ![]() ![]() ![]() |
SampledData | |||
![]() ![]() ![]() ![]() ![]() ![]() |
time | |||
![]() ![]() ![]() ![]() ![]() ![]() |
dateTime | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Period | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Attachment | |||
![]() ![]() ![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||
![]() ![]() ![]() ![]() |
SΣC | 1..1 | BackboneElement | The overall system connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-status | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Value Set (required) |
![]() ![]() ![]() ![]() |
SΣC | 0..* | BackboneElement | Additional modifiers to connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-modifier | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Modifier Value Set (required) |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing reasons why the patient may be listed at a given status and/or modifier. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-reason | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing steps to get the patient to a connected status |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-guidance | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-cied-connectivity-time | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-monitor-connectivity-time | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-cied-connectivity-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-monitor-connectivity-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the most recent remote device interrogation with a personal monitor. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-interrogation-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-interrogation-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-secondary-interrogation-date | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time interval in days between expected CIED and remote monitoring software communication. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connection-interval | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result |
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time period in days in which successful communication must take place for the connectivity status to remain connected. |
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized |
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: expected-time-period | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result |
![]() |
Path | Conformance | ValueSet / Code | URI |
Observation.status | required | ObservationStatushttp://hl7.org/fhir/ValueSet/observation-status|5.0.0 From the FHIR Standard | |
Observation.code | example | Pattern Value: observation-cied-connectivityhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component.code | example | LOINCCodes(a valid code from LOINC)http://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityStatus.code | example | Pattern Value: connectivity-statushttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityStatus.value[x] | required | CIEDConnectivityStateVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-vs From this IG | |
Observation.component:connectivityModifier.code | example | Pattern Value: connectivity-modifierhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityModifier.value[x] | required | CIEDConnectivityStateModifierVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-modifier-vs From this IG | |
Observation.component:statusReason.code | example | Pattern Value: status-reasonhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:statusGuidance.code | example | Pattern Value: status-guidancehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.code | example | Pattern Value: last-cied-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.code | example | Pattern Value: last-monitor-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.code | example | Pattern Value: next-cied-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.code | example | Pattern Value: next-monitor-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.code | example | Pattern Value: last-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.code | example | Pattern Value: next-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.code | example | Pattern Value: next-secondary-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectionInterval.code | example | Pattern Value: connection-intervalhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:expectedTimePeriod.code | example | Pattern Value: expected-time-periodhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard |
Id | Grade | Path(s) | Details | Requirements |
dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources : contained.contained.empty() | |
dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource : contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty() | |
dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated : contained.meta.versionId.empty() and contained.meta.lastUpdated.empty() | |
dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label : contained.meta.security.empty() | |
dom-6 | best practice | Observation | A resource should have narrative for robust management : text.`div`.exists() | |
ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children : hasValue() or (children().count() > id.count()) | |
ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both : extension.exists() != value.exists() | |
obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present : dataAbsentReason.empty() or value.empty() | |
obs-7 | error | Observation | If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). : value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty() | |
obs-8 | error | Observation | bodyStructure SHALL only be present if Observation.bodySite is not present : bodySite.exists() implies bodyStructure.empty() |
Differential View
This structure is derived from Observation
Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
---|---|---|---|---|
![]() ![]() |
0..* | Observation | Measurements and simple assertions | |
![]() ![]() ![]() |
1..* | Extension | Extension Slice: Unordered, Open by value:url | |
![]() ![]() ![]() ![]() |
S | 1..* | Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) | Cardiac implantable device(s) URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension |
![]() ![]() ![]() |
S | 1..1 | code | registered | preliminary | final | amended + |
![]() ![]() ![]() |
S | 1..1 | CodeableConcept | Type of observation (code / type) Required Pattern: At least the following |
![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: observation-cied-connectivity | |
![]() ![]() ![]() |
S | 1..1 | Reference(Patient - CIED Patient Profile) | Who and/or what the observation is about |
![]() ![]() ![]() |
S | 1..1 | dateTime | Clinically relevant time/time-period for observation |
![]() ![]() ![]() |
S | 0..1 | Reference(Device - CIED Monitor Profile) | A reference to the device that generates the measurements or the device settings for the device |
![]() ![]() ![]() |
S | 1..* | BackboneElement | CIED Connectivity Components Slice: Unordered, Open by value:code |
![]() ![]() ![]() ![]() |
Content/Rules for all slices | |||
![]() ![]() ![]() ![]() ![]() |
S | 1..1 | CodeableConcept | Type of component observation (code / type) |
![]() ![]() ![]() ![]() ![]() |
S | 0..1 | Quantity, CodeableConcept, string, boolean, integer, Range, Ratio, SampledData, time, dateTime, Period, Attachment, Reference(MolecularSequence) | Actual component result |
![]() ![]() ![]() ![]() |
S | 1..1 | BackboneElement | The overall system connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-status | |
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Value Set (required) | |
![]() ![]() ![]() ![]() |
S | 0..* | BackboneElement | Additional modifiers to connectivity status as calculated by the CIED manufacturer. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-modifier | |
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Modifier Value Set (required) | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | Text describing reasons why the patient may be listed at a given status and/or modifier. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-reason | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | Text describing steps to get the patient to a connected status |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-guidance | |
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the most recent communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-cied-connectivity-time | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the most recent communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-monitor-connectivity-time | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the next expected communication between the CIED and monitor. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-cied-connectivity-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The timestamp of the next expected communication from the monitor to the remote monitoring platform. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-monitor-connectivity-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The date of the most recent remote device interrogation with a personal monitor. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-interrogation-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-interrogation-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-secondary-interrogation-date | |
![]() ![]() ![]() ![]() ![]() |
0..1 | dateTime | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The time interval in days between expected CIED and remote monitoring software communication. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connection-interval | |
![]() ![]() ![]() ![]() ![]() |
0..1 | integer | Actual component result | |
![]() ![]() ![]() ![]() |
S | 0..1 | BackboneElement | The time period in days in which successful communication must take place for the connectivity status to remain connected. |
![]() ![]() ![]() ![]() ![]() |
1..1 | CodeableConcept | Type of component observation (code / type) Required Pattern: At least the following | |
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: expected-time-period | |
![]() ![]() ![]() ![]() ![]() |
0..1 | integer | Actual component result | |
![]() |
Path | Conformance | ValueSet | URI |
Observation.component:connectivityStatus.value[x] | required | CIEDConnectivityStateVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-vs From this IG | |
Observation.component:connectivityModifier.value[x] | required | CIEDConnectivityStateModifierVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-modifier-vs From this IG |
Snapshot View
Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() | ||||
---|---|---|---|---|---|---|---|---|
![]() ![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7, obs-8 | ||||
![]() ![]() ![]() |
Σ | 0..1 | id | Logical id of this artifact | ||||
![]() ![]() ![]() |
Σ | 0..1 | Meta | Metadata about the resource | ||||
![]() ![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created | ||||
![]() ![]() ![]() |
0..1 | code | Language of the resource content Binding: AllLanguages (required): IETF language tag for a human language
| |||||
![]() ![]() ![]() |
0..1 | Narrative | Text summary of the resource, for human interpretation This profile does not constrain the narrative in regard to content, language, or traceability to data elements | |||||
![]() ![]() ![]() |
0..* | Resource | Contained, inline Resources | |||||
![]() ![]() ![]() |
1..* | Extension | Extension Slice: Unordered, Open by value:url | |||||
![]() ![]() ![]() ![]() |
S | 1..* | Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) | Cardiac implantable device(s) URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension | ||||
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored | ||||
![]() ![]() ![]() |
Σ | 0..* | Identifier | Business Identifier for observation | ||||
![]() ![]() ![]() |
Σ | 0..1 | canonical(ObservationDefinition), Reference(ObservationDefinition) | Instantiates FHIR ObservationDefinition | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(CarePlan | DeviceRequest | ImmunizationRecommendation | MedicationRequest | NutritionOrder | ServiceRequest) | Fulfills plan, proposal or order | ||||
![]() ![]() ![]() |
0..* | BackboneElement | Triggering observation(s) | |||||
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() |
Σ | 1..1 | Reference(Observation) | Triggering observation | ||||
![]() ![]() ![]() ![]() |
Σ | 1..1 | code | reflex | repeat | re-run Binding: TriggeredBytype (required): The type of TriggeredBy Observation. | ||||
![]() ![]() ![]() ![]() |
0..1 | string | Reason that the observation was triggered | |||||
![]() ![]() ![]() |
Σ | 0..* | Reference(MedicationAdministration | MedicationDispense | MedicationStatement | Procedure | Immunization | ImagingStudy | GenomicStudy) | Part of referenced event | ||||
![]() ![]() ![]() |
?!SΣ | 1..1 | code | registered | preliminary | final | amended + Binding: ObservationStatus (required): Codes providing the status of an observation. | ||||
![]() ![]() ![]() |
0..* | CodeableConcept | Classification of type of observation Binding: ObservationCategoryCodes (preferred): Codes for high level observation categories. | |||||
![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: observation-cied-connectivity | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() |
SΣ | 1..1 | Reference(Patient - CIED Patient Profile) | Who and/or what the observation is about | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(Resource) | What the observation is about, when it is not about the subject of record | ||||
![]() ![]() ![]() |
Σ | 0..1 | Reference(Encounter) | Healthcare event during which this observation is made | ||||
![]() ![]() ![]() |
SΣ | 1..1 | dateTime | Clinically relevant time/time-period for observation | ||||
![]() ![]() ![]() |
Σ | 0..1 | instant | Date/Time this version was made available | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(Practitioner | PractitionerRole | Organization | CareTeam | Patient | RelatedPerson) | Who is responsible for the observation | ||||
![]() ![]() ![]() |
ΣC | 0..1 | Actual result | |||||
![]() ![]() ![]() ![]() |
Quantity | |||||||
![]() ![]() ![]() ![]() |
CodeableConcept | |||||||
![]() ![]() ![]() ![]() |
string | |||||||
![]() ![]() ![]() ![]() |
boolean | |||||||
![]() ![]() ![]() ![]() |
integer | |||||||
![]() ![]() ![]() ![]() |
Range | |||||||
![]() ![]() ![]() ![]() |
Ratio | |||||||
![]() ![]() ![]() ![]() |
SampledData | |||||||
![]() ![]() ![]() ![]() |
time | |||||||
![]() ![]() ![]() ![]() |
dateTime | |||||||
![]() ![]() ![]() ![]() |
Period | |||||||
![]() ![]() ![]() ![]() |
Attachment | |||||||
![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||||||
![]() ![]() ![]() |
C | 0..1 | CodeableConcept | Why the result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | ||||
![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() |
0..* | Annotation | Comments about the observation | |||||
![]() ![]() ![]() |
C | 0..1 | CodeableConcept | Observed body part Binding: SNOMEDCTBodyStructures (example): SNOMED CT Body site concepts | ||||
![]() ![]() ![]() |
C | 0..1 | Reference(BodyStructure) | Observed body structure | ||||
![]() ![]() ![]() |
0..1 | CodeableConcept | How it was done Binding: ObservationMethods (example): Methods for simple observations. | |||||
![]() ![]() ![]() |
C | 0..1 | Reference(Specimen | Group) | Specimen used for this observation Constraints: obs-9 | ||||
![]() ![]() ![]() |
S | 0..1 | Reference(Device - CIED Monitor Profile) | A reference to the device that generates the measurements or the device settings for the device | ||||
![]() ![]() ![]() |
C | 0..* | BackboneElement | Provides guide for interpretation Constraints: obs-3 | ||||
![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() |
C | 0..1 | SimpleQuantity | Low Range, if relevant | ||||
![]() ![]() ![]() ![]() |
C | 0..1 | SimpleQuantity | High Range, if relevant | ||||
![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Normal value, if relevant Binding: ObservationReferenceRangeNormalValueCodes (extensible): Codes identifying the normal value of the observation. | |||||
![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Reference range qualifier Binding: ObservationReferenceRangeMeaningCodes (preferred): Code for the meaning of a reference range. | |||||
![]() ![]() ![]() ![]() |
0..* | CodeableConcept | Reference range population Binding: ObservationReferenceRangeAppliesToCodes (example): Codes identifying the population the reference range applies to. | |||||
![]() ![]() ![]() ![]() |
0..1 | Range | Applicable age range, if relevant | |||||
![]() ![]() ![]() ![]() |
C | 0..1 | markdown | Text based reference range in an observation | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(Observation | QuestionnaireResponse | MolecularSequence) | Related resource that belongs to the Observation group | ||||
![]() ![]() ![]() |
Σ | 0..* | Reference(DocumentReference | ImagingStudy | ImagingSelection | QuestionnaireResponse | Observation | MolecularSequence | GenomicStudy) | Related resource from which the observation is made | ||||
![]() ![]() ![]() |
SΣC | 1..* | BackboneElement | CIED Connectivity Components Slice: Unordered, Open by value:code | ||||
![]() ![]() ![]() ![]() |
Content/Rules for all slices | |||||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. | ||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | Actual component result | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
Quantity | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
CodeableConcept | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
string | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
boolean | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
integer | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Range | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Ratio | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
SampledData | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
time | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
dateTime | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Period | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Attachment | |||||||
![]() ![]() ![]() ![]() ![]() ![]() |
Reference(MolecularSequence) | |||||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 1..1 | BackboneElement | The overall system connectivity status as calculated by the CIED manufacturer. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-status | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Value Set (required) | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..* | BackboneElement | Additional modifiers to connectivity status as calculated by the CIED manufacturer. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connectivity-modifier | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | CodeableConcept | Actual component result Binding: CIED Connectivity Status Modifier Value Set (required) | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing reasons why the patient may be listed at a given status and/or modifier. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-reason | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | Text describing steps to get the patient to a connected status | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: status-guidance | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | string | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication between the CIED and monitor. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-cied-connectivity-time | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the most recent communication from the monitor to the remote monitoring platform. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-monitor-connectivity-time | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication between the CIED and monitor. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-cied-connectivity-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The timestamp of the next expected communication from the monitor to the remote monitoring platform. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-monitor-connectivity-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the most recent remote device interrogation with a personal monitor. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: last-interrogation-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-interrogation-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: next-secondary-interrogation-date | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | dateTime | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time interval in days between expected CIED and remote monitoring software communication. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: connection-interval | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() ![]() ![]() ![]() |
SΣC | 0..1 | BackboneElement | The time period in days in which successful communication must take place for the connectivity status to remain connected. | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() ![]() ![]() |
SΣC | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. Required Pattern: At least the following | ||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
1..* | Coding | Code defined by a terminology system Fixed Value: (Complex) | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | uri | Identity of the terminology system Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CIEDconnectivity | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Version of the system - if relevant | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
1..1 | code | Symbol in syntax defined by the system Fixed Value: expected-time-period | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Representation defined by the system | |||||
![]() ![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | boolean | If this coding was chosen directly by the user | |||||
![]() ![]() ![]() ![]() ![]() ![]() |
0..1 | string | Plain text representation of the concept | |||||
![]() ![]() ![]() ![]() ![]() |
SΣ | 0..1 | integer | Actual component result | ||||
![]() ![]() ![]() ![]() ![]() |
0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
![]() |
Path | Conformance | ValueSet / Code | URI |
Observation.language | required | AllLanguageshttp://hl7.org/fhir/ValueSet/all-languages|5.0.0 From the FHIR Standard | |
Observation.triggeredBy.type | required | TriggeredBytypehttp://hl7.org/fhir/ValueSet/observation-triggeredbytype|5.0.0 From the FHIR Standard | |
Observation.status | required | ObservationStatushttp://hl7.org/fhir/ValueSet/observation-status|5.0.0 From the FHIR Standard | |
Observation.category | preferred | ObservationCategoryCodeshttp://hl7.org/fhir/ValueSet/observation-category From the FHIR Standard | |
Observation.code | example | Pattern Value: observation-cied-connectivityhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.bodySite | example | SNOMEDCTBodyStructureshttp://hl7.org/fhir/ValueSet/body-site From the FHIR Standard | |
Observation.method | example | ObservationMethodshttp://hl7.org/fhir/ValueSet/observation-methods From the FHIR Standard | |
Observation.referenceRange.normalValue | extensible | ObservationReferenceRangeNormalValueCodeshttp://hl7.org/fhir/ValueSet/observation-referencerange-normalvalue From the FHIR Standard | |
Observation.referenceRange.type | preferred | ObservationReferenceRangeMeaningCodeshttp://hl7.org/fhir/ValueSet/referencerange-meaning From the FHIR Standard | |
Observation.referenceRange.appliesTo | example | ObservationReferenceRangeAppliesToCodeshttp://hl7.org/fhir/ValueSet/referencerange-appliesto From the FHIR Standard | |
Observation.component.code | example | LOINCCodes(a valid code from LOINC)http://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:connectivityStatus.code | example | Pattern Value: connectivity-statushttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityStatus.value[x] | required | CIEDConnectivityStateVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-vs From this IG | |
Observation.component:connectivityStatus.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:connectivityStatus.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:connectivityModifier.code | example | Pattern Value: connectivity-modifierhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectivityModifier.value[x] | required | CIEDConnectivityStateModifierVShttp://hl7.org/fhir/uv/cardx-cied/ValueSet/cied-connectivity-state-modifier-vs From this IG | |
Observation.component:connectivityModifier.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:connectivityModifier.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:statusReason.code | example | Pattern Value: status-reasonhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:statusReason.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:statusReason.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:statusGuidance.code | example | Pattern Value: status-guidancehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:statusGuidance.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:statusGuidance.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.code | example | Pattern Value: last-cied-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:lastCiedConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.code | example | Pattern Value: last-monitor-connectivity-timehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:lastMonitorConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.code | example | Pattern Value: next-cied-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextCiedConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.code | example | Pattern Value: next-monitor-connectivity-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextMonitorConnectivityDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.code | example | Pattern Value: last-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:lastRemoteInterrogationDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.code | example | Pattern Value: next-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextScheduledRemoteInterrogationDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.code | example | Pattern Value: next-secondary-interrogation-datehttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:connectionInterval.code | example | Pattern Value: connection-intervalhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:connectionInterval.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:connectionInterval.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard | |
Observation.component:expectedTimePeriod.code | example | Pattern Value: expected-time-periodhttp://hl7.org/fhir/ValueSet/observation-codes From the FHIR Standard | |
Observation.component:expectedTimePeriod.dataAbsentReason | extensible | DataAbsentReasonhttp://hl7.org/fhir/ValueSet/data-absent-reason From the FHIR Standard | |
Observation.component:expectedTimePeriod.interpretation | extensible | ObservationInterpretationCodeshttp://hl7.org/fhir/ValueSet/observation-interpretation From the FHIR Standard |
Id | Grade | Path(s) | Details | Requirements |
dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources : contained.contained.empty() | |
dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource : contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty() | |
dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated : contained.meta.versionId.empty() and contained.meta.lastUpdated.empty() | |
dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label : contained.meta.security.empty() | |
dom-6 | best practice | Observation | A resource should have narrative for robust management : text.`div`.exists() | |
ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children : hasValue() or (children().count() > id.count()) | |
ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both : extension.exists() != value.exists() | |
obs-3 | error | Observation.referenceRange | Must have at least a low or a high or text : low.exists() or high.exists() or text.exists() | |
obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present : dataAbsentReason.empty() or value.empty() | |
obs-7 | error | Observation | If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). : value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty() | |
obs-8 | error | Observation | bodyStructure SHALL only be present if Observation.bodySite is not present : bodySite.exists() implies bodyStructure.empty() | |
obs-9 | error | Observation.specimen | If Observation.specimen is a reference to Group, the group can only have specimens : (reference.resolve().exists() and reference.resolve() is Group) implies reference.resolve().member.entity.resolve().all($this is Specimen) |
This structure is derived from Observation
Summary
Mandatory: 6 elements
Must-Support: 22 elements
Structures
This structure refers to these other structures:
Extensions
This structure refers to these extensions:
Slices
This structure defines the following Slices:
Maturity: 2
Other representations of profile: CSV, Excel, Schematron