This page is part of the Personal Health Device FHIR IG (v0.3.0: STU 1 Ballot 3) based on FHIR R4. The current version which supercedes this version is 1.0.0. For a full list of available versions, see the Directory of published versions
Format(s):
Narrative view
Generated Narrative with Details
id: bits-1.0.0.40
meta:
status: final
code: MDC_PULS_OXIM_DEV_STATUS: Measuring process issues (Details : {urn:iso:std:iso:11073:10101 code '150604' = '150604)
effective: 12/11/2018 11:07:48 AM
component
code: sensor-displaced (Details : {http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 code '150604.2' = 'Sensor incorrectly placed on user)
value: Sensor is incorrectly placed on user (Details : {http://terminology.hl7.org/CodeSystem/v2-0136 code 'Y' = 'Yes)
component
code: signal-pulse-questionable (Details : {http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 code '150604.7' = 'Signal pulse questionable)
value: Questionable pulse detected (Details : {http://terminology.hl7.org/CodeSystem/v2-0136 code 'Y' = 'Yes)
component
code: signal-low-perfusion (Details : {http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 code '150604.10' = 'Signal low perfusion)
value: Signal experiencing low perfusion (Details : {http://terminology.hl7.org/CodeSystem/v2-0136 code 'Y' = 'Yes)
component
code: signal-poor (Details : {http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 code '150604.11' = 'Signal is poor)
value: Signal is poor (Details : {http://terminology.hl7.org/CodeSystem/v2-0136 code 'Y' = 'Yes)
component
code: signal-inadequate (Details : {http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 code '150604.12' = 'Signal inadequate)
value: Signal is inadequate (Details : {http://terminology.hl7.org/CodeSystem/v2-0136 code 'Y' = 'Yes)