This page is part of the FHIR Specification v4.3.0-snapshot1: R4B Snapshot to support the Jan 2022 Connectathon. About the R4B version of FHIR. The current officially released version is 4.3.0. For a full list of available versions, see the Directory of published versions . Page versions: R5 R4B R4 R3 R2
Health Care Devices Work Group | Maturity Level: 1 | Trial Use | Security Category: Anonymous | Compartments: Not linked to any defined compartments |
Detailed Descriptions for the elements in the DeviceMetric resource.
DeviceMetric | |
Element Id | DeviceMetric |
Definition | Describes a measurement, calculation or setting capability of a medical device. |
Cardinality | 0..* |
Type | DomainResource |
Summary | false |
Comments | For the initial scope, this DeviceMetric resource is only applicable to describe a single metric node in the containment tree that is produced by the context scanner in any medical device that implements or derives from the ISO/IEEE 11073 standard. |
DeviceMetric.identifier | |
Element Id | DeviceMetric.identifier |
Definition | Unique instance identifiers assigned to a device by the device or gateway software, manufacturers, other organizations or owners. For example: handle ID. |
Note | This is a business identifier, not a resource identifier (see discussion) |
Cardinality | 0..* |
Type | Identifier |
Summary | true |
Comments | For identifiers assigned to a device by the device or gateway software, the |
DeviceMetric.type | |
Element Id | DeviceMetric.type |
Definition | Describes the type of the metric. For example: Heart Rate, PEEP Setting, etc. |
Cardinality | 1..1 |
Terminology Binding | Device Metric and Component Types (Preferred) |
Type | CodeableConcept |
Summary | true |
Comments | DeviceMetric.type can be referred to either IEEE 11073-10101 or LOINC. |
DeviceMetric.unit | |
Element Id | DeviceMetric.unit |
Definition | Describes the unit that an observed value determined for this metric will have. For example: Percent, Seconds, etc. |
Cardinality | 0..1 |
Terminology Binding | Device Metric and Component Types (Preferred) |
Type | CodeableConcept |
Summary | true |
Comments | DeviceMetric.unit can refer to either UCUM or preferable a RTMMS coding system. |
DeviceMetric.source | |
Element Id | DeviceMetric.source |
Definition | Describes the link to the Device that this DeviceMetric belongs to and that contains administrative device information such as manufacturer, serial number, etc. |
Cardinality | 0..1 |
Type | Reference(Device) |
Summary | true |
DeviceMetric.parent | |
Element Id | DeviceMetric.parent |
Definition | Describes the link to the Device that this DeviceMetric belongs to and that provide information about the location of this DeviceMetric in the containment structure of the parent Device. An example would be a Device that represents a Channel. This reference can be used by a client application to distinguish DeviceMetrics that have the same type, but should be interpreted based on their containment location. |
Cardinality | 0..1 |
Type | Reference(Device) |
Summary | true |
DeviceMetric.operationalStatus | |
Element Id | DeviceMetric.operationalStatus |
Definition | Indicates current operational state of the device. For example: On, Off, Standby, etc. |
Cardinality | 0..1 |
Terminology Binding | DeviceMetricOperationalStatus (Required) |
Type | code |
Summary | true |
DeviceMetric.color | |
Element Id | DeviceMetric.color |
Definition | Describes the color representation for the metric. This is often used to aid clinicians to track and identify parameter types by color. In practice, consider a Patient Monitor that has ECG/HR and Pleth for example; the parameters are displayed in different characteristic colors, such as HR-blue, BP-green, and PR and SpO2- magenta. |
Cardinality | 0..1 |
Terminology Binding | DeviceMetricColor (Required) |
Type | code |
Summary | true |
DeviceMetric.category | |
Element Id | DeviceMetric.category |
Definition | Indicates the category of the observation generation process. A DeviceMetric can be for example a setting, measurement, or calculation. |
Cardinality | 1..1 |
Terminology Binding | DeviceMetricCategory (Required) |
Type | code |
Summary | true |
DeviceMetric.measurementPeriod | |
Element Id | DeviceMetric.measurementPeriod |
Definition | Describes the measurement repetition time. This is not necessarily the same as the update period. The measurement repetition time can range from milliseconds up to hours. An example for a measurement repetition time in the range of milliseconds is the sampling rate of an ECG. An example for a measurement repetition time in the range of hours is a NIBP that is triggered automatically every hour. The update period may be different than the measurement repetition time, if the device does not update the published observed value with the same frequency as it was measured. |
Cardinality | 0..1 |
Type | Timing |
Summary | true |
DeviceMetric.calibration | |
Element Id | DeviceMetric.calibration |
Definition | Describes the calibrations that have been performed or that are required to be performed. |
Cardinality | 0..* |
Summary | true |
DeviceMetric.calibration.type | |
Element Id | DeviceMetric.calibration.type |
Definition | Describes the type of the calibration method. |
Cardinality | 0..1 |
Terminology Binding | DeviceMetricCalibrationType (Required) |
Type | code |
Summary | true |
DeviceMetric.calibration.state | |
Element Id | DeviceMetric.calibration.state |
Definition | Describes the state of the calibration. |
Cardinality | 0..1 |
Terminology Binding | DeviceMetricCalibrationState (Required) |
Type | code |
Summary | true |
DeviceMetric.calibration.time | |
Element Id | DeviceMetric.calibration.time |
Definition | Describes the time last calibration has been performed. |
Cardinality | 0..1 |
Type | instant |
Summary | true |