This page is part of the Point-of-care Device FHIR IG (v0.1.0: STU 1 Draft) based on FHIR R3. . For a full list of available versions, see the Directory of published versions

PoC Device Profile

Canonical URL

http://devices.fhir.org/StructureDefinition/PoCDevice

Profile Definition

NameFlagsCard.Type
extension0..*Extension
modifierExtension?!0..*Extension
identifier0..*Identifier
extension0..*Extension
useΣ ?!0..1codeBinding
extension0..*Extension
extension0..*Extension
systemΣ1..1uriFixed Value
versionΣ0..1string
codeΣ1..1codeFixed Value
displayΣ0..1string
userSelectedΣ0..1boolean
textΣ0..1string
systemΣ0..1uri
valueΣ1..1string
periodΣ0..1Period
assignerΣ0..1Reference(Organization)
extension0..*Extension
useΣ ?!0..1codeBinding
typeΣ0..1CodeableConceptBinding
systemΣ1..1uriFixed Value
valueΣ1..1string
periodΣ0..1Period
assignerΣ0..1Reference(Organization)
extension0..*Extension
deviceIdentifierΣ0..1string
nameΣ0..1string
jurisdiction0..1uri
carrierHRFΣ0..1string
carrierAIDCΣ0..1base64Binary
issuer0..1uri
entryType0..1codeBinding
statusΣ ?!0..1codeBinding
extension0..*Extension
codingΣ1..*Coding
extension0..*Extension
systemΣ1..1uriFixed Value
versionΣ0..1string
codeΣ1..1code
displayΣ0..1string
userSelectedΣ0..1boolean
textΣ0..1string
lotNumber0..1string
manufacturer0..1string
manufactureDate0..1dateTime
expirationDate0..1dateTime
model0..1string
version0..1string
patient0..1Reference(Patient)
owner0..1Reference(Organization)
contact0..*ContactPoint
location0..1Reference(Location)
url0..1uri
note0..*Annotation
safetyΣ0..*CodeableConcept

Example (XML view)

Example of a Physio Monitor Device

<Device>
    <id value="72" />
    <meta>
        <profile value="http://devices.fhir.org/StructureDefinition/PoCDevice" />
    </meta>
    <identifier>
        <type>
            <coding>
                <system value="http://hl7.org/fhir/identifier-type" />
                <code value="SNO" />
            </coding>
        </type>
        <value value="DE35115712" />
    </identifier>
    <identifier>
        <system value="urn:oid:1.2.840.10004.1.1.1.0.0.1.0.0.1.2680" />
        <value value="00-09-FB-FF-FF-A0-F4-CA" />
    </identifier>
    <udi>
        <deviceIdentifier value="00884838038752" />
        <jurisdiction value="http://hl7.org/fhir/NamingSystem/fda-udi" />
        <carrierHRF value="(01)00884838038752(21)DE35115712" />
        <issuer value="http://hl7.org/fhir/NamingSystem/gs1-di" />
    </udi>
    <status value="active" />
    <type>
        <coding>
            <system value="urn:iso:std:iso:11073:10101" />
            <code value="69965" />
            <display value="MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS" />
        </coding>
        <text value="Patient Monitor" />
    </type>
    <manufacturer value="Philips" />
    <model value="MX400 - MX550" />
    <patient>
        <reference value="Patient/371" />
    </patient>
    <location>
        <reference value="Location/46" />
    </location>
</Device>

Example (JSON view)

Example of a Physio Monitor Device

{
    "resourceType":"Device",
    "id":"72",
    "meta":{
        "profile": [
            "http://devices.fhir.org/StructureDefinition/PoCDevice"
        ]
    },
    "identifier": [
        {
            "type":{
                "coding": [
                    {
                        "system":"http://hl7.org/fhir/identifier-type",
                        "code":"SNO"
                    }
                ]
            },
            "value":"DE35115712"
        },
        {
            "system":"urn:oid:1.2.840.10004.1.1.1.0.0.1.0.0.1.2680",
            "value":"00-09-FB-FF-FF-A0-F4-CA"
        }
    ],
    "udi":{
        "deviceIdentifier":"00884838038752",
        "jurisdiction":"http://hl7.org/fhir/NamingSystem/fda-udi",
        "carrierHRF":"(01)00884838038752(21)DE35115712",
        "issuer":"http://hl7.org/fhir/NamingSystem/gs1-di"
    },
    "status":"active",
    "type":{
        "coding": [
            {
                "system":"urn:iso:std:iso:11073:10101",
                "code":"69965",
                "display":"MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS"
            }
        ],
        "text":"Patient Monitor"
    },
    "manufacturer":"Philips",
    "model":"MX400 - MX550",
    "patient":{
        "reference":"Patient/371"
    },
    "location":{
        "reference":"Location/46"
    }
}