This page is part of the Point-of-care Device FHIR IG (v0.3.0: STU 1 Ballot 3) based on FHIR R4. . For a full list of available versions, see the Directory of published versions
Definitions for the EnumerationDeviceMetric resource profile.
1. DeviceMetric | |
Definition | Describes a measurement, calculation or setting capability of a medical device. |
Control | 0..* |
Comments | For the initial scope, this DeviceMetric resource is only applicable to describe a single metric node in the containment tree that is produced by the context scanner in any medical device that implements or derives from the ISO/IEEE 11073 standard. |
Invariants | Defined on this element dom-2: If the resource is contained in another resource, it SHALL NOT contain nested Resources (: contained.contained.empty()) dom-3: If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource (: contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()) dom-4: If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated (: contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()) dom-5: If a resource is contained in another resource, it SHALL NOT have a security label (: contained.meta.security.empty()) dom-6: A resource should have narrative for robust management (: text.`div`.exists()) |
2. DeviceMetric.id | |
Definition | The logical id of the resource, as used in the URL for the resource. Once assigned, this value never changes. |
Control | 0..1 |
Type | string |
Comments | The only time that a resource does not have an id is when it is being submitted to the server using a create operation. |
3. DeviceMetric.meta | |
Definition | The metadata about the resource. This is content that is maintained by the infrastructure. Changes to the content might not always be associated with version changes to the resource. |
Control | 0..1 |
Type | Meta |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
4. DeviceMetric.implicitRules | |
Definition | A reference to a set of rules that were followed when the resource was constructed, and which must be understood when processing the content. Often, this is a reference to an implementation guide that defines the special rules along with other profiles etc. |
Control | 0..1 |
Type | uri |
Is Modifier | true |
Comments | Asserting this rule set restricts the content to be only understood by a limited set of trading partners. This inherently limits the usefulness of the data in the long term. However, the existing health eco-system is highly fractured, and not yet ready to define, collect, and exchange data in a generally computable sense. Wherever possible, implementers and/or specification writers should avoid using this element. Often, when used, the URL is a reference to an implementation guide that defines these special rules as part of it's narrative along with other profiles, value sets, etc. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
5. DeviceMetric.language | |
Definition | The base language in which the resource is written. |
Control | 0..1 |
Binding | The codes SHOULD be taken from CommonLanguages Max Binding: AllLanguages A human language. |
Type | code |
Comments | Language is provided to support indexing and accessibility (typically, services such as text to speech use the language tag). The html language tag in the narrative applies to the narrative. The language tag on the resource may be used to specify the language of other presentations generated from the data in the resource. Not all the content has to be in the base language. The Resource.language should not be assumed to apply to the narrative automatically. If a language is specified, it should it also be specified on the div element in the html (see rules in HTML5 for information about the relationship between xml:lang and the html lang attribute). |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
6. DeviceMetric.text | |
Definition | A human-readable narrative that contains a summary of the resource and can be used to represent the content of the resource to a human. The narrative need not encode all the structured data, but is required to contain sufficient detail to make it "clinically safe" for a human to just read the narrative. Resource definitions may define what content should be represented in the narrative to ensure clinical safety. |
Control | 0..1 |
Type | Narrative |
Alternate Names | narrative, html, xhtml, display |
Comments | Contained resources do not have narrative. Resources that are not contained SHOULD have a narrative. In some cases, a resource may only have text with little or no additional discrete data (as long as all minOccurs=1 elements are satisfied). This may be necessary for data from legacy systems where information is captured as a "text blob" or where text is additionally entered raw or narrated and encoded information is added later. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
7. DeviceMetric.contained | |
Definition | These resources do not have an independent existence apart from the resource that contains them - they cannot be identified independently, and nor can they have their own independent transaction scope. |
Control | 0..* |
Type | Resource |
Alternate Names | inline resources, anonymous resources, contained resources |
Comments | This should never be done when the content can be identified properly, as once identification is lost, it is extremely difficult (and context dependent) to restore it again. Contained resources may have profiles and tags In their meta elements, but SHALL NOT have security labels. |
8. DeviceMetric.extension | |
Definition | An Extension |
Control | 0..* |
Type | Extension |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
9. DeviceMetric.extension:instance | |
SliceName | instance |
Definition | An Extension |
Control | 0..1 This element is affected by the following invariants: ele-1 |
Type | Extension(DeviceInstance) (Extension Type: Choice of: base64Binary, boolean, canonical, code, date, dateTime, decimal, id, instant, integer, markdown, oid, positiveInt, string, time, unsignedInt, uri, url, uuid, Address, Age, Annotation, Attachment, CodeableConcept, Coding, ContactPoint, Count, Distance, Duration, HumanName, Identifier, Money, Period, Quantity, Range, Ratio, Reference, SampledData, Signature, Timing, ContactDetail, Contributor, DataRequirement, Expression, ParameterDefinition, RelatedArtifact, TriggerDefinition, UsageContext, Dosage, Meta) |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
10. DeviceMetric.extension:instance.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
11. DeviceMetric.extension:instance.extension | |
Definition | An Extension |
Control | 0..* |
Type | Extension |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.extension.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
12. DeviceMetric.extension:instance.extension:number | |
SliceName | number |
Definition | Instance number for multiple components of the same type |
Control | 0..1 |
Type | Extension |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
13. DeviceMetric.extension:instance.extension:number.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
14. DeviceMetric.extension:instance.extension:number.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.extension.extension.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
15. DeviceMetric.extension:instance.extension:number.extension.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
16. DeviceMetric.extension:instance.extension:number.extension.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.extension.extension.extension.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
17. DeviceMetric.extension:instance.extension:number.extension.url | |
Definition | Source of the definition for the extension code - a logical name or a URL. |
Control | 1..1 |
Type | uri |
Comments | The definition may point directly to a computable or human-readable definition of the extensibility codes, or it may be a logical URI as declared in some other specification. The definition SHALL be a URI for the Structure Definition defining the extension. |
18. DeviceMetric.extension:instance.extension:number.extension.value[x] | |
Definition | Value of extension - must be one of a constrained set of the data types (see Extensibility for a list). |
Control | 0..1 |
Type | Choice of: base64Binary, boolean, canonical, code, date, dateTime, decimal, id, instant, integer, markdown, oid, positiveInt, string, time, unsignedInt, uri, url, uuid, Address, Age, Annotation, Attachment, CodeableConcept, Coding, ContactPoint, Count, Distance, Duration, HumanName, Identifier, Money, Period, Quantity, Range, Ratio, Reference, SampledData, Signature, Timing, ContactDetail, Contributor, DataRequirement, Expression, ParameterDefinition, RelatedArtifact, TriggerDefinition, UsageContext, Dosage, Meta |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
19. DeviceMetric.extension:instance.extension:number.url | |
Definition | Source of the definition for the extension code - a logical name or a URL. |
Control | 1..1 |
Type | uri |
Comments | The definition may point directly to a computable or human-readable definition of the extensibility codes, or it may be a logical URI as declared in some other specification. The definition SHALL be a URI for the Structure Definition defining the extension. |
Fixed Value | number |
20. DeviceMetric.extension:instance.extension:number.value[x] | |
Definition | Value of extension - must be one of a constrained set of the data types (see Extensibility for a list). |
Control | 1..1 |
Type | unsignedInt |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
Slicing | This element introduces a set of slices on DeviceMetric.extension.extension.value[x]. The slices are unordered and Closed, and can be differentiated using the following discriminators:
|
21. DeviceMetric.extension:instance.extension:number.value[x]:valueUnsignedInt | |
SliceName | valueUnsignedInt |
Definition | Value of extension - must be one of a constrained set of the data types (see Extensibility for a list). |
Control | 1..1 |
Type | unsignedInt |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
22. DeviceMetric.extension:instance.extension:label | |
SliceName | label |
Definition | An Extension |
Control | 0..1 |
Type | Extension |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
23. DeviceMetric.extension:instance.extension:label.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
24. DeviceMetric.extension:instance.extension:label.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.extension.extension.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
25. DeviceMetric.extension:instance.extension:label.extension.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
26. DeviceMetric.extension:instance.extension:label.extension.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.extension.extension.extension.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
27. DeviceMetric.extension:instance.extension:label.extension.url | |
Definition | Source of the definition for the extension code - a logical name or a URL. |
Control | 1..1 |
Type | uri |
Comments | The definition may point directly to a computable or human-readable definition of the extensibility codes, or it may be a logical URI as declared in some other specification. The definition SHALL be a URI for the Structure Definition defining the extension. |
28. DeviceMetric.extension:instance.extension:label.extension.value[x] | |
Definition | Value of extension - must be one of a constrained set of the data types (see Extensibility for a list). |
Control | 0..1 |
Type | Choice of: base64Binary, boolean, canonical, code, date, dateTime, decimal, id, instant, integer, markdown, oid, positiveInt, string, time, unsignedInt, uri, url, uuid, Address, Age, Annotation, Attachment, CodeableConcept, Coding, ContactPoint, Count, Distance, Duration, HumanName, Identifier, Money, Period, Quantity, Range, Ratio, Reference, SampledData, Signature, Timing, ContactDetail, Contributor, DataRequirement, Expression, ParameterDefinition, RelatedArtifact, TriggerDefinition, UsageContext, Dosage, Meta |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
29. DeviceMetric.extension:instance.extension:label.url | |
Definition | Source of the definition for the extension code - a logical name or a URL. |
Control | 1..1 |
Type | uri |
Comments | The definition may point directly to a computable or human-readable definition of the extensibility codes, or it may be a logical URI as declared in some other specification. The definition SHALL be a URI for the Structure Definition defining the extension. |
Fixed Value | label |
30. DeviceMetric.extension:instance.extension:label.value[x] | |
Definition | Value of extension - must be one of a constrained set of the data types (see Extensibility for a list). |
Control | 1..1 |
Type | string |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
Slicing | This element introduces a set of slices on DeviceMetric.extension.extension.value[x]. The slices are unordered and Closed, and can be differentiated using the following discriminators:
|
31. DeviceMetric.extension:instance.extension:label.value[x]:valueString | |
SliceName | valueString |
Definition | Value of extension - must be one of a constrained set of the data types (see Extensibility for a list). |
Control | 1..1 |
Type | string |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
32. DeviceMetric.extension:instance.url | |
Definition | Source of the definition for the extension code - a logical name or a URL. |
Control | 1..1 |
Type | uri |
Comments | The definition may point directly to a computable or human-readable definition of the extensibility codes, or it may be a logical URI as declared in some other specification. The definition SHALL be a URI for the Structure Definition defining the extension. |
Fixed Value | http://hl7.org/fhir/uv/pocd/StructureDefinition/device-instance |
33. DeviceMetric.extension:instance.value[x] | |
Definition | Value of extension - must be one of a constrained set of the data types (see Extensibility for a list). |
Control | 0..0 |
Type | Choice of: base64Binary, boolean, canonical, code, date, dateTime, decimal, id, instant, integer, markdown, oid, positiveInt, string, time, unsignedInt, uri, url, uuid, Address, Age, Annotation, Attachment, CodeableConcept, Coding, ContactPoint, Count, Distance, Duration, HumanName, Identifier, Money, Period, Quantity, Range, Ratio, Reference, SampledData, Signature, Timing, ContactDetail, Contributor, DataRequirement, Expression, ParameterDefinition, RelatedArtifact, TriggerDefinition, UsageContext, Dosage, Meta |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
34. DeviceMetric.extension:relation | |
SliceName | relation |
Definition | Relation allows the modelling of relationships between a metric and other containment tree entries. Related containment tree entries are defined in Entries, whereby the flavor of a relationship can be set up in Kind. |
Control | 0..* This element is affected by the following invariants: ele-1 |
Type | Extension(Relation) (Extension Type: Choice of: base64Binary, boolean, canonical, code, date, dateTime, decimal, id, instant, integer, markdown, oid, positiveInt, string, time, unsignedInt, uri, url, uuid, Address, Age, Annotation, Attachment, CodeableConcept, Coding, ContactPoint, Count, Distance, Duration, HumanName, Identifier, Money, Period, Quantity, Range, Ratio, Reference, SampledData, Signature, Timing, ContactDetail, Contributor, DataRequirement, Expression, ParameterDefinition, RelatedArtifact, TriggerDefinition, UsageContext, Dosage, Meta) |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
35. DeviceMetric.extension:metricAvailability | |
SliceName | metricAvailability |
Definition | Availability of the means that derives the metric state. |
Control | 0..1 This element is affected by the following invariants: ele-1 |
Type | Extension(MetricAvailability) (Extension Type: code) |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
36. DeviceMetric.modifierExtension | |
Definition | May be used to represent additional information that is not part of the basic definition of the resource and that modifies the understanding of the element that contains it and/or the understanding of the containing element's descendants. Usually modifier elements provide negation or qualification. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer is allowed to define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. Applications processing a resource are required to check for modifier extensions. Modifier extensions SHALL NOT change the meaning of any elements on Resource or DomainResource (including cannot change the meaning of modifierExtension itself). |
Control | 0..* |
Type | Extension |
Is Modifier | true |
Requirements | Modifier extensions allow for extensions that cannot be safely ignored to be clearly distinguished from the vast majority of extensions which can be safely ignored. This promotes interoperability by eliminating the need for implementers to prohibit the presence of extensions. For further information, see the definition of modifier extensions. |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
37. DeviceMetric.identifier | |
Definition | Unique instance identifiers assigned to a device by the device or gateway software, manufacturers, other organizations or owners. For example: handle ID. |
Note | This is a business identifier, not a resource identifier (see discussion) |
Control | 0..* |
Type | Identifier |
Comments | For identifiers assigned to a device by the device or gateway software, the |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
38. DeviceMetric.type | |
Definition | Describes the type of the metric. For example: Heart Rate, PEEP Setting, etc. |
Control | 1..1 |
Binding | The codes SHALL be taken from MDC Metric nomenclature; other codes may be used where these codes are not suitable |
Type | CodeableConcept |
Must Support | true |
Comments | DeviceMetric.type can be referred to either IEEE 11073-10101 or LOINC. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
39. DeviceMetric.type.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
40. DeviceMetric.type.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.type.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
41. DeviceMetric.type.coding | |
Definition | A reference to a code defined by a terminology system. |
Control | 0..* |
Type | Coding |
Requirements | Allows for alternative encodings within a code system, and translations to other code systems. |
Comments | Codes may be defined very casually in enumerations, or code lists, up to very formal definitions such as SNOMED CT - see the HL7 v3 Core Principles for more information. Ordering of codings is undefined and SHALL NOT be used to infer meaning. Generally, at most only one of the coding values will be labeled as UserSelected = true. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
42. DeviceMetric.type.coding.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
43. DeviceMetric.type.coding.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.type.coding.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
44. DeviceMetric.type.coding.system | |
Definition | The identification of the code system that defines the meaning of the symbol in the code. |
Control | 0..1 |
Type | uri |
Requirements | Need to be unambiguous about the source of the definition of the symbol. |
Comments | The URI may be an OID (urn:oid:...) or a UUID (urn:uuid:...). OIDs and UUIDs SHALL be references to the HL7 OID registry. Otherwise, the URI should come from HL7's list of FHIR defined special URIs or it should reference to some definition that establishes the system clearly and unambiguously. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
45. DeviceMetric.type.coding.version | |
Definition | The version of the code system which was used when choosing this code. Note that a well-maintained code system does not need the version reported, because the meaning of codes is consistent across versions. However this cannot consistently be assured, and when the meaning is not guaranteed to be consistent, the version SHOULD be exchanged. |
Note | This is a business versionId, not a resource version id (see discussion) |
Control | 0..1 |
Type | string |
Comments | Where the terminology does not clearly define what string should be used to identify code system versions, the recommendation is to use the date (expressed in FHIR date format) on which that version was officially published as the version date. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
46. DeviceMetric.type.coding.code | |
Definition | A symbol in syntax defined by the system. The symbol may be a predefined code or an expression in a syntax defined by the coding system (e.g. post-coordination). |
Control | 0..1 |
Type | code |
Requirements | Need to refer to a particular code in the system. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
47. DeviceMetric.type.coding.display | |
Definition | A representation of the meaning of the code in the system, following the rules of the system. |
Control | 0..1 |
Type | string |
Requirements | Need to be able to carry a human-readable meaning of the code for readers that do not know the system. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
48. DeviceMetric.type.coding.userSelected | |
Definition | Indicates that this coding was chosen by a user directly - e.g. off a pick list of available items (codes or displays). |
Control | 0..1 |
Type | boolean |
Requirements | This has been identified as a clinical safety criterium - that this exact system/code pair was chosen explicitly, rather than inferred by the system based on some rules or language processing. |
Comments | Amongst a set of alternatives, a directly chosen code is the most appropriate starting point for new translations. There is some ambiguity about what exactly 'directly chosen' implies, and trading partner agreement may be needed to clarify the use of this element and its consequences more completely. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
49. DeviceMetric.type.text | |
Definition | A human language representation of the concept as seen/selected/uttered by the user who entered the data and/or which represents the intended meaning of the user. |
Control | 0..1 |
Type | string |
Requirements | The codes from the terminologies do not always capture the correct meaning with all the nuances of the human using them, or sometimes there is no appropriate code at all. In these cases, the text is used to capture the full meaning of the source. |
Comments | Very often the text is the same as a displayName of one of the codings. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
50. DeviceMetric.unit | |
Definition | Describes the unit that an observed value determined for this metric will have. For example: Percent, Seconds, etc. |
Control | 0..0 |
Binding | The codes SHOULD be taken from DeviceMetricAndComponentTypes Describes the unit of the metric. |
Type | CodeableConcept |
Comments | DeviceMetric.unit can refer to either UCUM or preferable a RTMMS coding system. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
51. DeviceMetric.unit.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
52. DeviceMetric.unit.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.unit.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
53. DeviceMetric.unit.coding | |
Definition | A reference to a code defined by a terminology system. |
Control | 0..* |
Type | Coding |
Requirements | Allows for alternative encodings within a code system, and translations to other code systems. |
Comments | Codes may be defined very casually in enumerations, or code lists, up to very formal definitions such as SNOMED CT - see the HL7 v3 Core Principles for more information. Ordering of codings is undefined and SHALL NOT be used to infer meaning. Generally, at most only one of the coding values will be labeled as UserSelected = true. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
54. DeviceMetric.unit.coding.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
55. DeviceMetric.unit.coding.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.unit.coding.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
56. DeviceMetric.unit.coding.system | |
Definition | The identification of the code system that defines the meaning of the symbol in the code. |
Control | 0..1 |
Type | uri |
Requirements | Need to be unambiguous about the source of the definition of the symbol. |
Comments | The URI may be an OID (urn:oid:...) or a UUID (urn:uuid:...). OIDs and UUIDs SHALL be references to the HL7 OID registry. Otherwise, the URI should come from HL7's list of FHIR defined special URIs or it should reference to some definition that establishes the system clearly and unambiguously. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
57. DeviceMetric.unit.coding.version | |
Definition | The version of the code system which was used when choosing this code. Note that a well-maintained code system does not need the version reported, because the meaning of codes is consistent across versions. However this cannot consistently be assured, and when the meaning is not guaranteed to be consistent, the version SHOULD be exchanged. |
Note | This is a business versionId, not a resource version id (see discussion) |
Control | 0..1 |
Type | string |
Comments | Where the terminology does not clearly define what string should be used to identify code system versions, the recommendation is to use the date (expressed in FHIR date format) on which that version was officially published as the version date. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
58. DeviceMetric.unit.coding.code | |
Definition | A symbol in syntax defined by the system. The symbol may be a predefined code or an expression in a syntax defined by the coding system (e.g. post-coordination). |
Control | 0..1 |
Type | code |
Requirements | Need to refer to a particular code in the system. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
59. DeviceMetric.unit.coding.display | |
Definition | A representation of the meaning of the code in the system, following the rules of the system. |
Control | 0..1 |
Type | string |
Requirements | Need to be able to carry a human-readable meaning of the code for readers that do not know the system. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
60. DeviceMetric.unit.coding.userSelected | |
Definition | Indicates that this coding was chosen by a user directly - e.g. off a pick list of available items (codes or displays). |
Control | 0..1 |
Type | boolean |
Requirements | This has been identified as a clinical safety criterium - that this exact system/code pair was chosen explicitly, rather than inferred by the system based on some rules or language processing. |
Comments | Amongst a set of alternatives, a directly chosen code is the most appropriate starting point for new translations. There is some ambiguity about what exactly 'directly chosen' implies, and trading partner agreement may be needed to clarify the use of this element and its consequences more completely. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
61. DeviceMetric.unit.text | |
Definition | A human language representation of the concept as seen/selected/uttered by the user who entered the data and/or which represents the intended meaning of the user. |
Control | 0..1 |
Type | string |
Requirements | The codes from the terminologies do not always capture the correct meaning with all the nuances of the human using them, or sometimes there is no appropriate code at all. In these cases, the text is used to capture the full meaning of the source. |
Comments | Very often the text is the same as a displayName of one of the codings. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
62. DeviceMetric.source | |
Definition | Describes the link to the Device that this DeviceMetric belongs to and that contains administrative device information such as manufacturer, serial number, etc. |
Control | 1..1 |
Type | Reference(MdsDevice) |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
63. DeviceMetric.parent | |
Definition | Describes the link to the Device that this DeviceMetric belongs to and that provide information about the location of this DeviceMetric in the containment structure of the parent Device. An example would be a Device that represents a Channel. This reference can be used by a client application to distinguish DeviceMetrics that have the same type, but should be interpreted based on their containment location. |
Control | 1..1 |
Type | Reference(ChannelDevice) |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
64. DeviceMetric.operationalStatus | |
Definition | Indicates current operational state of the device. For example: On, Off, Standby, etc. |
Control | 0..1 |
Binding | The codes SHALL be taken from DeviceMetricOperationalStatus Describes the operational status of the DeviceMetric. |
Type | code |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
65. DeviceMetric.color | |
Definition | Describes the color representation for the metric. This is often used to aid clinicians to track and identify parameter types by color. In practice, consider a Patient Monitor that has ECG/HR and Pleth for example; the parameters are displayed in different characteristic colors, such as HR-blue, BP-green, and PR and SpO2- magenta. |
Control | 0..1 |
Binding | The codes SHALL be taken from DeviceMetricColor Describes the typical color of representation. |
Type | code |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
66. DeviceMetric.category | |
Definition | Indicates the category of the observation generation process. A DeviceMetric can be for example a setting, measurement, or calculation. |
Control | 1..1 |
Binding | The codes SHALL be taken from DeviceMetricCategory Describes the category of the metric. |
Type | code |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
67. DeviceMetric.measurementPeriod | |
Definition | Describes the measurement repetition time. This is not necessarily the same as the update period. The measurement repetition time can range from milliseconds up to hours. An example for a measurement repetition time in the range of milliseconds is the sampling rate of an ECG. An example for a measurement repetition time in the range of hours is a NIBP that is triggered automatically every hour. The update period may be different than the measurement repetition time, if the device does not update the published observed value with the same frequency as it was measured. |
Control | 0..1 |
Type | Timing |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
68. DeviceMetric.measurementPeriod.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
69. DeviceMetric.measurementPeriod.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.measurementPeriod.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
70. DeviceMetric.measurementPeriod.modifierExtension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element and that modifies the understanding of the element in which it is contained and/or the understanding of the containing element's descendants. Usually modifier elements provide negation or qualification. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. Applications processing a resource are required to check for modifier extensions. Modifier extensions SHALL NOT change the meaning of any elements on Resource or DomainResource (including cannot change the meaning of modifierExtension itself). |
Control | 0..* |
Type | Extension |
Is Modifier | true |
Requirements | Modifier extensions allow for extensions that cannot be safely ignored to be clearly distinguished from the vast majority of extensions which can be safely ignored. This promotes interoperability by eliminating the need for implementers to prohibit the presence of extensions. For further information, see the definition of modifier extensions. |
Alternate Names | extensions, user content, modifiers |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
71. DeviceMetric.measurementPeriod.event | |
Definition | Identifies specific times when the event occurs. |
Control | 0..* |
Type | dateTime |
Requirements | In a Medication Administration Record, for instance, you need to take a general specification, and turn it into a precise specification. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
72. DeviceMetric.measurementPeriod.repeat | |
Definition | A set of rules that describe when the event is scheduled. |
Control | 0..1 |
Type | Element |
Requirements | Many timing schedules are determined by regular repetitions. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) tim-1: if there's a duration, there needs to be duration units (: duration.empty() or durationUnit.exists()) tim-10: If there's a timeOfDay, there cannot be a when, or vice versa (: timeOfDay.empty() or when.empty()) tim-2: if there's a period, there needs to be period units (: period.empty() or periodUnit.exists()) tim-4: duration SHALL be a non-negative value (: duration.exists() implies duration >= 0) tim-5: period SHALL be a non-negative value (: period.exists() implies period >= 0) tim-6: If there's a periodMax, there must be a period (: periodMax.empty() or period.exists()) tim-7: If there's a durationMax, there must be a duration (: durationMax.empty() or duration.exists()) tim-8: If there's a countMax, there must be a count (: countMax.empty() or count.exists()) tim-9: If there's an offset, there must be a when (and not C, CM, CD, CV) (: offset.empty() or (when.exists() and ((when in ('C' | 'CM' | 'CD' | 'CV')).not()))) |
73. DeviceMetric.measurementPeriod.repeat.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
74. DeviceMetric.measurementPeriod.repeat.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
Slicing | This element introduces a set of slices on DeviceMetric.measurementPeriod.repeat.extension. The slices are unordered and Open, and can be differentiated using the following discriminators:
|
75. DeviceMetric.measurementPeriod.repeat.bounds[x] | |
Definition | Either a duration for the length of the timing schedule, a range of possible length, or outer bounds for start and/or end limits of the timing schedule. |
Control | 0..1 |
Type | Choice of: Duration, Range, Period |
[x] Note | See Choice of Data Types for further information about how to use [x] |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
76. DeviceMetric.measurementPeriod.repeat.count | |
Definition | A total count of the desired number of repetitions across the duration of the entire timing specification. If countMax is present, this element indicates the lower bound of the allowed range of count values. |
Control | 0..1 |
Type | positiveInt |
Requirements | Repetitions may be limited by end time or total occurrences. |
Comments | If you have both bounds and count, then this should be understood as within the bounds period, until count times happens. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
77. DeviceMetric.measurementPeriod.repeat.countMax | |
Definition | If present, indicates that the count is a range - so to perform the action between [count] and [countMax] times. |
Control | 0..1 |
Type | positiveInt |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
78. DeviceMetric.measurementPeriod.repeat.duration | |
Definition | How long this thing happens for when it happens. If durationMax is present, this element indicates the lower bound of the allowed range of the duration. |
Control | 0..1 |
Type | decimal |
Requirements | Some activities are not instantaneous and need to be maintained for a period of time. |
Comments | For some events the duration is part of the definition of the event (e.g. IV infusions, where the duration is implicit in the specified quantity and rate). For others, it's part of the timing specification (e.g. exercise). |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
79. DeviceMetric.measurementPeriod.repeat.durationMax | |
Definition | If present, indicates that the duration is a range - so to perform the action between [duration] and [durationMax] time length. |
Control | 0..1 |
Type | decimal |
Requirements | Some activities are not instantaneous and need to be maintained for a period of time. |
Comments | For some events the duration is part of the definition of the event (e.g. IV infusions, where the duration is implicit in the specified quantity and rate). For others, it's part of the timing specification (e.g. exercise). |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
80. DeviceMetric.measurementPeriod.repeat.durationUnit | |
Definition | The units of time for the duration, in UCUM units. |
Control | 0..1 |
Binding | The codes SHALL be taken from UnitsOfTime A unit of time (units from UCUM). |
Type | code |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
81. DeviceMetric.measurementPeriod.repeat.frequency | |
Definition | The number of times to repeat the action within the specified period. If frequencyMax is present, this element indicates the lower bound of the allowed range of the frequency. |
Control | 0..1 |
Type | positiveInt |
Meaning if Missing | If no frequency is stated, the assumption is that the event occurs once per period, but systems SHOULD always be specific about this |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
82. DeviceMetric.measurementPeriod.repeat.frequencyMax | |
Definition | If present, indicates that the frequency is a range - so to repeat between [frequency] and [frequencyMax] times within the period or period range. |
Control | 0..1 |
Type | positiveInt |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
83. DeviceMetric.measurementPeriod.repeat.period | |
Definition | Indicates the duration of time over which repetitions are to occur; e.g. to express "3 times per day", 3 would be the frequency and "1 day" would be the period. If periodMax is present, this element indicates the lower bound of the allowed range of the period length. |
Control | 0..1 |
Type | decimal |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
84. DeviceMetric.measurementPeriod.repeat.periodMax | |
Definition | If present, indicates that the period is a range from [period] to [periodMax], allowing expressing concepts such as "do this once every 3-5 days. |
Control | 0..1 |
Type | decimal |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
85. DeviceMetric.measurementPeriod.repeat.periodUnit | |
Definition | The units of time for the period in UCUM units. |
Control | 0..1 |
Binding | The codes SHALL be taken from UnitsOfTime A unit of time (units from UCUM). |
Type | code |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
86. DeviceMetric.measurementPeriod.repeat.dayOfWeek | |
Definition | If one or more days of week is provided, then the action happens only on the specified day(s). |
Control | 0..* |
Binding | The codes SHALL be taken from DaysOfWeek |
Type | code |
Comments | If no days are specified, the action is assumed to happen every day as otherwise specified. The elements frequency and period cannot be used as well as dayOfWeek. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
87. DeviceMetric.measurementPeriod.repeat.timeOfDay | |
Definition | Specified time of day for action to take place. |
Control | 0..* |
Type | time |
Comments | When time of day is specified, it is inferred that the action happens every day (as filtered by dayofWeek) on the specified times. The elements when, frequency and period cannot be used as well as timeOfDay. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
88. DeviceMetric.measurementPeriod.repeat.when | |
Definition | An approximate time period during the day, potentially linked to an event of daily living that indicates when the action should occur. |
Control | 0..* |
Binding | The codes SHALL be taken from EventTiming Real world event relating to the schedule. |
Type | code |
Requirements | Timings are frequently determined by occurrences such as waking, eating and sleep. |
Comments | When more than one event is listed, the event is tied to the union of the specified events. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
89. DeviceMetric.measurementPeriod.repeat.offset | |
Definition | The number of minutes from the event. If the event code does not indicate whether the minutes is before or after the event, then the offset is assumed to be after the event. |
Control | 0..1 |
Type | unsignedInt |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
90. DeviceMetric.measurementPeriod.code | |
Definition | A code for the timing schedule (or just text in code.text). Some codes such as BID are ubiquitous, but many institutions define their own additional codes. If a code is provided, the code is understood to be a complete statement of whatever is specified in the structured timing data, and either the code or the data may be used to interpret the Timing, with the exception that .repeat.bounds still applies over the code (and is not contained in the code). |
Control | 0..1 |
Binding | The codes SHOULD be taken from TimingAbbreviation Code for a known / defined timing pattern. |
Type | CodeableConcept |
Comments | BID etc. are defined as 'at institutionally specified times'. For example, an institution may choose that BID is "always at 7am and 6pm". If it is inappropriate for this choice to be made, the code BID should not be used. Instead, a distinct organization-specific code should be used in place of the HL7-defined BID code and/or a structured representation should be used (in this case, specifying the two event times). |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
91. DeviceMetric.calibration | |
Definition | Describes the calibrations that have been performed or that are required to be performed. |
Control | 0..* |
Type | BackboneElement |
Must Support | true |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
92. DeviceMetric.calibration.id | |
Definition | Unique id for the element within a resource (for internal references). This may be any string value that does not contain spaces. |
Control | 0..1 |
Type | string |
93. DeviceMetric.calibration.extension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. |
Control | 0..* |
Type | Extension |
Alternate Names | extensions, user content |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
94. DeviceMetric.calibration.modifierExtension | |
Definition | May be used to represent additional information that is not part of the basic definition of the element and that modifies the understanding of the element in which it is contained and/or the understanding of the containing element's descendants. Usually modifier elements provide negation or qualification. To make the use of extensions safe and manageable, there is a strict set of governance applied to the definition and use of extensions. Though any implementer can define an extension, there is a set of requirements that SHALL be met as part of the definition of the extension. Applications processing a resource are required to check for modifier extensions. Modifier extensions SHALL NOT change the meaning of any elements on Resource or DomainResource (including cannot change the meaning of modifierExtension itself). |
Control | 0..* |
Type | Extension |
Is Modifier | true |
Requirements | Modifier extensions allow for extensions that cannot be safely ignored to be clearly distinguished from the vast majority of extensions which can be safely ignored. This promotes interoperability by eliminating the need for implementers to prohibit the presence of extensions. For further information, see the definition of modifier extensions. |
Alternate Names | extensions, user content, modifiers |
Comments | There can be no stigma associated with the use of extensions by any application, project, or standard - regardless of the institution or jurisdiction that uses or defines the extensions. The use of extensions is what allows the FHIR specification to retain a core level of simplicity for everyone. |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) ext-1: Must have either extensions or value[x], not both (: extension.exists() != value.exists()) |
95. DeviceMetric.calibration.type | |
Definition | Describes the type of the calibration method. |
Control | 0..1 |
Binding | The codes SHALL be taken from DeviceMetricCalibrationType Describes the type of a metric calibration. |
Type | code |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
96. DeviceMetric.calibration.state | |
Definition | Describes the state of the calibration. |
Control | 0..1 |
Binding | The codes SHALL be taken from DeviceMetricCalibrationState Describes the state of a metric calibration. |
Type | code |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |
97. DeviceMetric.calibration.time | |
Definition | Describes the time last calibration has been performed. |
Control | 0..1 |
Type | instant |
Invariants | Defined on this element ele-1: All FHIR elements must have a @value or children (: hasValue() or (children().count() > id.count())) |