CodeSystem: ASN1ToHL7

Official URL: http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 Version: 1.1.0
Active as of 2024-10-18 Computable Name: ASN1ToHL7

Maps IEEE 11073 ASN1 BITS measurements to codes. All these codes have two possible values, set or cleared. The code may either represent an event or a state. When an event, the defined event has occurred when set. If a state, the described item is either set or cleared. Sometimes the states may be defined in unexpected manners. For example, the regulation state when set is unregulated and when cleared it is regulated. In addition the code may derive from an IEEE 11073 attribute or from a measurement. More codes maybe added to this list in the future.

This Code system is referenced in the content logical definition of the following value sets:

Generated Narrative: CodeSystem ASN1ToHL7

Properties

This code system defines the following properties for its concepts

NameCodeURITypeDescription
type type https://sagroups.ieee.org/11073/TypeIsASN1EventOrState code Indicates that the code type is an event or state
source source https://sagroups.ieee.org/11073/SourceIsMeasurementOrDevice string Indicates that the code originates from a measurement or an attribute that further describes a measurement or the sensor device.

Concepts

This case-sensitive code system http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 defines the following codes in a Grouped By hierarchy:

LvlCodeDisplayDefinition
1 67925.0 onMains Indicates whether the PHD is on or off mains power
1 67925.1 onBattery Indicates whether the PHD is on or off Battery power
1 67925.8 chargingFull Indicates whether the PHD battery is charging at maximum rate
1 67925.9 chargingTrickle Indicates whether the PHD battery is charging at a trickle rate
1 67925.10 chargingOff Indicates whether the PHD battery charging is on or off
1 532354.0 regulation-status Indicates device is regulated when cleared
1 68219.0 mds-time-capab-real-time-clock Indicates real time clock is supported
1 68219.1 mds-time-capab-set-clock Indicates support for the PHG to set the time
1 68219.2 mds-time-capab-relative-time Indicates Relative Time is supported
1 68219.3 mds-time-capab-high-res-relative-time Indicates Hi-Res Relative Time is supported
1 68219.4 mds-time-capab-sync-abs-time Indicates that syncing the absolute time is supported
1 68219.5 mds-time-capab-sync-rel-time Indicates that syncing the relative time is supported
1 68219.6 mds-time-capab-sync-hi-res-relative-time Indicates that syncing hi-res relative time is supported
1 68219.7 mds-time-capab-bo-time Indicates that the real-time clock supports base offset time
1 68219.8 mds-time-state-abs-time-synced Indicates the absolute time has been synced
1 68219.9 mds-time-state-rel-time-synced Indicates the relative time has been synced
1 68219.10 mds-time-state-hi-res-relative-time-synced Indicates the hi-res time has been synced
1 68219.11 mds-time-mgr-set-time PHG Requested to set the time
1 68219.12 mds-time-capab-sync-bo-time Indicates that syncing base-offset time is supported
1 68219.13 mds-time-state-bo-time-synced Indicates that base-offset time has been synced
1 68219.14 mds-time-state-bo-time-UTC-aligned Indicates that base-offset time is aligned with UTC time standard
1 68219.15 mds-time-dst-rules-enabled Indicates support for daylight savings time rules
1 67846.0 lim-alert-off Indicates both the high and the low limit alerts are disabled
1 67846.1 lim-low-off Indicates the low limit alert is disabled
1 67846.2 lim-high-off Indicates the high limit alert is disabled
1 150604.0 sensor-disconnected Indicates the device is not connected with the sensor
1 150604.1 sensor-malfunction Indicates the sensor has malfunctioned
1 150604.2 sensor-displaced Indicates the sensor is not correctly placed on the user
1 150604.3 sensor-unsupported Indicates the sensor connected to the device is unsupported
1 150604.4 sensor-off Indicates the sensor is not connected to the user
1 150604.5 sensor-interference Indicates the sensor is experiencing interference
1 150604.6 signal-searching Indicates the sensor is searching for a signal
1 150604.7 signal-pulse-questionable Indicates a questionable pulse has been detected
1 150604.8 signal-non-pulsatile Indicates a non pulsatile signal has been detected
1 150604.9 signal-erratic Indicates a signal has been discovered, but it is erratic
1 150604.10 signal-low-perfusion Indicates the signal is experiencing low perfusion
1 150604.11 signal-poor Indicates the signal has been discovered, but it is poor
1 150604.12 signal-inadequate Indicates the signal has been discovered, but it is inadequate
1 150604.13 signal-processing-irregularity Indicates the signal has been discovered, but is experiencing processing irregularity
1 150604.14 device-equipment-malfunction Indicates a device malfunction has occurred
1 150604.15 device-extended-update Indicates an extended display update is in progress on device
1 150605.0 pulse-qual-nominal Indicates no abnormalities have been detected by the sensor
1 150605.1 pulse-qual-marginal Indicates the pulse profusion or quality is marginal
1 150605.2 pulse-qual-minimal Indicates the pulse profusion or quality is minimal
1 150605.3 pulse-qual-unacceptable Indicates the pulse profusion or quality is unacceptable
1 8410584.0 leadwire-loss Loss of unspecified lead wire or electrode connection
1 8410584.1 leadsignal-loss Loss of unspecified lead signal
1 8410584.2 leadwire-loss-first-lead Loss of 1st lead wire or electrode connection
1 8410584.3 leadsignal-loss-first-lead Loss of 1st lead signal
1 8410584.4 leadwire-loss-second-lead Loss of 2nd lead wire or electrode connection
1 8410584.5 leadsignal-loss-second-lead Loss of 2nd lead signal
1 8410584.6 leadwire-loss-third-lead Loss of 3rd lead wire or electrode connection
1 8410584.7 leadsignal-loss-third-lead Loss of 3rd lead signal
1 8417752.0 device-battery-low The device battery is low
1 8417752.1 sensor-malfunction The sensor has malfunctioned
1 8417752.2 sensor-sample-size-insufficient The sensor's sample size is insufficient
1 8417752.3 sensor-strip-insertion The test strip was inserted incorrectly
1 8417752.4 sensor-strip-type-incorrect The test strip type is incompatable with the sensor
1 8417752.5 sensor-result-too-high The value reported by the sensor is too high
1 8417752.6 sensor-result-too-low The value reported by the sensor is too low
1 8417752.7 sensor-temp-too-high The temperature is too high
1 8417752.8 sensor-temp-too-low The temperature is too low
1 8417752.9 sensor-read-interrupt The read process was interrupted
1 8417752.10 device-gen-fault General Device Fault
1 8417752.11 sensor-temp-out-of-range The temperature is out of range
1 8417909.0 inr-device-battery-low The INR device battery is low
1 8417909.1 inr-sensor-malfunction The INR device sensor has malfunctioned
1 8417909.2 inr-sensor-sample-size-insufficient The sample size was insufficient
1 8417909.3 inr-sensor-strip-insertion The test script was inserted incorrectly
1 8417909.4 inr-sensor-strip-type-incorrect The test strip used is not compatible with the INR
1 8417909.5 inr-sensor-result-too-high The test result is too high
1 8417909.6 inr-sensor-result-too-low The test result is too low
1 8417909.7 inr-sensor-temp-too-high The temperature is too high
1 8417909.8 inr-sensor-temp-too-low The temperature is too low
1 8417909.9 inr-sensor-read-interrupt The read process was interrupted
1 8417909.10 inr-device-gen-fault General device fault
1 8417909.11 inr-sensor-calibration-due The INR sensor calibration is due
1 8408608.0 device-status-undetermined Undetermined device status event
1 8408608.1 device-status-reset Device reset
1 8408608.5 device-status-error Device error
1 8408608.6 device-status-error-mechanical Mechanical error
1 8408608.7 device-status-error-electronic Electronic error
1 8408608.8 device-status-error-software Software error
1 8408608.9 device-status-error-battery Battery error
1 8408608.15 device-status-service Undetermined device status event, service needed
1 8408608.16 device-status-service-time-sync-required Time sync service required
1 8408608.17 device-status-service-calibration-required Calibration required
1 8408608.18 device-status-service-replenishment-required Replenishment required
1 8408608.25 device-status-battery-low Battery power is too low
1 8408608.26 device-status-battery-depleted Battery is depleted
1 8408608.27 device-status-battery-replaced Battery has been replaced
1 8408608.28 device-status-battery-interrupted Battery was interrupted
1 8418060.0 sensor-session-stopped Session stopped
1 8418060.2 sensor-type-incorrect Sensor type incorrect for device
1 8418060.3 sensor-malfunction The sensor malfunctioned
1 8418060.4 device-specific-alert Device specific alert
1 8418060.7 sensor-calibration-not-allowed Calibrating the sensor is not allowed
1 8418060.8 sensor-calibration-recommended Calibrating the sensor is recommended
1 8418060.9 sensor-calibration-required Calibrating the sensor is required
1 8418060.10 sensor-temp-too-high Temperature is too high for a valid result
1 8418060.11 sensor-temp-too-low Temperature is too low for a valid result
1 8418060.12 sensor-result-below-patient-low Sensor result below the Patient Low level
1 8418060.13 sensor-result-above-patient-high Sensor result above the Patient High level
1 8418060.14 sensor-low-hypo Sensor result below level
1 8418060.15 sensor-high-hyper Sensor result above level
1 8418060.16 sensor-rate-decrease-exceeded The sensor rate of decrease has been exceeded
1 8418060.17 sensor-rate-increase-exceeded The sensor rate of increase has been exceeded
1 8418060.18 sensor-result-too-low Sensor result is lower than the device can process
1 8418060.19 sensor-result-too-high Sensor result is higher than the device can process
1 8418060.20 sensor-com-out-of-range The sensor communication is out of range
1 8418512.0 Battery-status-Undetermined the status of the battery is unknown
1 8418512.1 Battery-absent Indicates if the battery is absent
1 8418512.2 Battery-active Indicates if the battery is active in the system
1 8418512.3 Battery-charging Indicates if the battery is charging
1 8418512.4 Battery-fullyCharged Indicates if the battery is fully charged
1 8418512.5 Battery-disposable Indicates if the battery is disposable
1 8418512.6 Battery-rechargeable Indicates if battery is rechargable
1 8418512.8 Battery-faulty Indicates if battery is faulty
2   8418512.7 Battery-overTemperature Indicates if the battery is too hot
1 8418512.9 Battery-incompatible Indicates if battery is incompatible
1 8410608.0 body-movement Indicates too much body movement
1 8410608.1 cuff-too-loose Indicates the cuff was too loose
1 8410608.2 irregular-pulse Indicates an irregular pulse was detected
1 8410608.3 pulse-over-range-limit Indicates the pulse was too high for sensor to work
1 8410608.4 pulse-under-range-limit Indicates the pulse was too low for sensor to work
1 8410608.5 improper-body-position Indicates the cuff was incorrectly placed with respect to the heart