This page is part of the Personal Health Device FHIR IG (v1.1.0: STU 1) based on FHIR (HL7® FHIR® Standard) R4. This is the current published version. For a full list of available versions, see the Directory of published versions
This example results when the device sends a waveform. The value[x] becomes a valueSampledData data type.
Note there is no logical id in this resource as this resource is being uploaded to the server in a create operation. The server will create the logical id and return it to the sender in the response.
{
"resourceType": "Observation",
"meta": {
"profile": [
"http://hl7.org/fhir/uv/phd/StructureDefinition/PhdRtsaObservation"
]
},
"category" : [
{
"coding" : [
{
"system" : "http://hl7.org/fhir/uv/phd/CodeSystem/PhdObservationCategories",
"code" : "phd-observation"
}
]
}
],
"extension": [
{
"url": "http://hl7.org/fhir/StructureDefinition/observation-gatewayDevice",
"valueReference": {
"reference": "Device/phg-ecde3d4e58532d31.000000000000"
}
}
],
"status": "final",
"code": {
"coding": [
{
"system": "urn:iso:std:iso:11073:10101",
"code": "150452"
},
{
"code": "SampledData",
"display": "Data is a periodic array"
}
],
"text": "MDC_PULS_OXIM_PLETH: Pleth Wave"
},
"subject": {
"reference": "Patient/sisansarahId.1.2.3.4.5.6.7.8.10"
},
"effectiveDateTime": "2018-08-02T02:25:24-04:00",
"valueSampledData": {
"origin": {
"value": 0,
"system": "http://unitsofmeasure.org",
"code": "{unitless} 1"
},
"period": 0.01,
"factor": 3.0,
"dimensions": 1,
"data": "123 110 97 99 112 118"
},
"device": {
"reference": "Device/phd-00601900010E9234.F45EABA80832"
},
"derivedFrom": [
{
"reference": "Observation/coin-1234"
}
]
}