This page is part of the Personal Health Device FHIR IG (v2.0.0-ballot: STU 2 Ballot 1) based on FHIR (HL7® FHIR® Standard) v4.3.0. The current version which supersedes this version is 1.0.0. For a full list of available versions, see the Directory of published versions
Every IEEE 11073-10206 Observation instance is required to have a Type attribute. The Type attribute value is an MDC term code that maps directly to the Observation.code element. The Observation.code element is populated as follows:
The requirement for the MDC textual reference identifier is not mandatory and is not used as the Observation.code.coding.code value because the reference identifier is not provided by the sensor device in the exchange protocol. Requiring the reference identifier would require that the PHG have an internal map. However, the main reason for not requiring the reference identifier is that doing so would defeat future interoperability. An older PHG would not know the reference identifier of a new specialization but it would be able to propagate the new code since that is provided by the device.
The MDC code entry is required, but the concept may be expressed in other coding systems as well. The choice of any additional translations is up to the application. PHGs operating in certain realms may be required to provide translations to a realm-specified coding system. For example, in the United States realm, LOINC is strongly preferred. For those implementing a PHG for operations in the US, a mapping table for MDC to LOINC can be downloaded here. One may freely use this table as needed in an implementation license free.
The consumer of FHIR resources containing the PHD data does not need to concern itself with this complexity. All it needs to do is obtain the Observation.code.coding.code element supporting the 11073-10101 system value to get the measurement type.