Lvl | Code | Display | Definition | Type of bit - indicates if a bit represents an Event (that happens) or a State (that persists) | Source - indicates if a bit comes from a measurement or from a device |
1 |
67925.0 |
onMains |
Indicates whether the PHD is on or off mains power |
state |
attribute |
1 |
67925.1 |
onBattery |
Indicates whether the PHD is on or off Battery power |
state |
attribute |
1 |
67925.8 |
chargingFull |
Indicates whether the PHD battery is charging at maximum rate |
event |
attribute |
1 |
67925.9 |
chargingTrickle |
Indicates whether the PHD battery is charging at a trickle rate |
event |
attribute |
1 |
67925.10 |
chargingOff |
Indicates whether the PHD battery charging is on or off |
state |
attribute |
1 |
532354.0 |
regulation-status |
Indicates device is regulated when cleared |
state |
attribute |
1 |
68219.0 |
mds-time-capab-real-time-clock |
Indicates real time clock is supported |
event |
attribute |
1 |
68219.1 |
mds-time-capab-set-clock |
Indicates support for the PHG to set the time |
event |
attribute |
1 |
68219.2 |
mds-time-capab-relative-time |
Indicates Relative Time is supported |
event |
attribute |
1 |
68219.3 |
mds-time-capab-high-res-relative-time |
Indicates Hi-Res Relative Time is supported |
event |
attribute |
1 |
68219.4 |
mds-time-capab-sync-abs-time |
Indicates that syncing the absolute time is supported |
event |
attribute |
1 |
68219.5 |
mds-time-capab-sync-rel-time |
Indicates that syncing the relative time is supported |
event |
attribute |
1 |
68219.6 |
mds-time-capab-sync-hi-res-relative-time |
Indicates that syncing hi-res relative time is supported |
event |
attribute |
1 |
68219.7 |
mds-time-capab-bo-time |
Indicates that the real-time clock supports base offset time |
event |
attribute |
1 |
68219.8 |
mds-time-state-abs-time-synced |
Indicates the absolute time has been synced |
event |
attribute |
1 |
68219.9 |
mds-time-state-rel-time-synced |
Indicates the relative time has been synced |
event |
attribute |
1 |
68219.10 |
mds-time-state-hi-res-relative-time-synced |
Indicates the hi-res time has been synced |
event |
attribute |
1 |
68219.11 |
mds-time-mgr-set-time |
PHG Requested to set the time |
event |
attribute |
1 |
68219.12 |
mds-time-capab-sync-bo-time |
Indicates that syncing base-offset time is supported |
event |
attribute |
1 |
68219.13 |
mds-time-state-bo-time-synced |
Indicates that base-offset time has been synced |
event |
attribute |
1 |
68219.14 |
mds-time-state-bo-time-UTC-aligned |
Indicates that base-offset time is aligned with UTC time standard |
event |
attribute |
1 |
68219.15 |
mds-time-dst-rules-enabled |
Indicates support for daylight savings time rules |
event |
attribute |
1 |
67846.0 |
lim-alert-off |
Indicates both the high and the low limit alerts are disabled |
state |
measurement |
1 |
67846.1 |
lim-low-off |
Indicates the low limit alert is disabled |
state |
measurement |
1 |
67846.2 |
lim-high-off |
Indicates the high limit alert is disabled |
state |
measurement |
1 |
150604.0 |
sensor-disconnected |
Indicates the device is not connected with the sensor |
event |
measurement |
1 |
150604.1 |
sensor-malfunction |
Indicates the sensor has malfunctioned |
event |
measurement |
1 |
150604.2 |
sensor-displaced |
Indicates the sensor is not correctly placed on the user |
event |
measurement |
1 |
150604.3 |
sensor-unsupported |
Indicates the sensor connected to the device is unsupported |
event |
measurement |
1 |
150604.4 |
sensor-off |
Indicates the sensor is not connected to the user |
event |
measurement |
1 |
150604.5 |
sensor-interference |
Indicates the sensor is experiencing interference |
event |
measurement |
1 |
150604.6 |
signal-searching |
Indicates the sensor is searching for a signal |
event |
measurement |
1 |
150604.7 |
signal-pulse-questionable |
Indicates a questionable pulse has been detected |
event |
measurement |
1 |
150604.8 |
signal-non-pulsatile |
Indicates a non pulsatile signal has been detected |
event |
measurement |
1 |
150604.9 |
signal-erratic |
Indicates a signal has been discovered, but it is erratic |
event |
measurement |
1 |
150604.10 |
signal-low-perfusion |
Indicates the signal is experiencing low perfusion |
event |
measurement |
1 |
150604.11 |
signal-poor |
Indicates the signal has been discovered, but it is poor |
event |
measurement |
1 |
150604.12 |
signal-inadequate |
Indicates the signal has been discovered, but it is inadequate |
event |
measurement |
1 |
150604.13 |
signal-processing-irregularity |
Indicates the signal has been discovered, but is experiencing processing irregularity |
event |
measurement |
1 |
150604.14 |
device-equipment-malfunction |
Indicates a device malfunction has occurred |
event |
measurement |
1 |
150604.15 |
device-extended-update |
Indicates an extended display update is in progress on device |
event |
measurement |
1 |
150605.0 |
pulse-qual-nominal |
Indicates no abnormalities have been detected by the sensor |
event |
measurement |
1 |
150605.1 |
pulse-qual-marginal |
Indicates the pulse profusion or quality is marginal |
event |
measurement |
1 |
150605.2 |
pulse-qual-minimal |
Indicates the pulse profusion or quality is minimal |
event |
measurement |
1 |
150605.3 |
pulse-qual-unacceptable |
Indicates the pulse profusion or quality is unacceptable |
event |
measurement |
1 |
8410584.0 |
leadwire-loss |
Loss of unspecified lead wire or electrode connection |
event |
measurement |
1 |
8410584.1 |
leadsignal-loss |
Loss of unspecified lead signal |
event |
measurement |
1 |
8410584.2 |
leadwire-loss-first-lead |
Loss of 1st lead wire or electrode connection |
event |
measurement |
1 |
8410584.3 |
leadsignal-loss-first-lead |
Loss of 1st lead signal |
event |
measurement |
1 |
8410584.4 |
leadwire-loss-second-lead |
Loss of 2nd lead wire or electrode connection |
event |
measurement |
1 |
8410584.5 |
leadsignal-loss-second-lead |
Loss of 2nd lead signal |
event |
measurement |
1 |
8410584.6 |
leadwire-loss-third-lead |
Loss of 3rd lead wire or electrode connection |
event |
measurement |
1 |
8410584.7 |
leadsignal-loss-third-lead |
Loss of 3rd lead signal |
event |
measurement |
1 |
8417752.0 |
device-battery-low |
The device battery is low |
event |
measurement |
1 |
8417752.1 |
sensor-malfunction |
The sensor has malfunctioned |
event |
measurement |
1 |
8417752.2 |
sensor-sample-size-insufficient |
The sensor's sample size is insufficient |
event |
measurement |
1 |
8417752.3 |
sensor-strip-insertion |
The test strip was inserted incorrectly |
event |
measurement |
1 |
8417752.4 |
sensor-strip-type-incorrect |
The test strip type is incompatable with the sensor |
event |
measurement |
1 |
8417752.5 |
sensor-result-too-high |
The value reported by the sensor is too high |
event |
measurement |
1 |
8417752.6 |
sensor-result-too-low |
The value reported by the sensor is too low |
event |
measurement |
1 |
8417752.7 |
sensor-temp-too-high |
The temperature is too high |
event |
measurement |
1 |
8417752.8 |
sensor-temp-too-low |
The temperature is too low |
event |
measurement |
1 |
8417752.9 |
sensor-read-interrupt |
The read process was interrupted |
event |
measurement |
1 |
8417752.10 |
device-gen-fault |
General Device Fault |
event |
measurement |
1 |
8417752.11 |
sensor-temp-out-of-range |
The temperature is out of range |
event |
measurement |
1 |
8417909.0 |
inr-device-battery-low |
The INR device battery is low |
event |
measurement |
1 |
8417909.1 |
inr-sensor-malfunction |
The INR device sensor has malfunctioned |
event |
measurement |
1 |
8417909.2 |
inr-sensor-sample-size-insufficient |
The sample size was insufficient |
event |
measurement |
1 |
8417909.3 |
inr-sensor-strip-insertion |
The test script was inserted incorrectly |
event |
measurement |
1 |
8417909.4 |
inr-sensor-strip-type-incorrect |
The test strip used is not compatible with the INR |
event |
measurement |
1 |
8417909.5 |
inr-sensor-result-too-high |
The test result is too high |
event |
measurement |
1 |
8417909.6 |
inr-sensor-result-too-low |
The test result is too low |
event |
measurement |
1 |
8417909.7 |
inr-sensor-temp-too-high |
The temperature is too high |
event |
measurement |
1 |
8417909.8 |
inr-sensor-temp-too-low |
The temperature is too low |
event |
measurement |
1 |
8417909.9 |
inr-sensor-read-interrupt |
The read process was interrupted |
event |
measurement |
1 |
8417909.10 |
inr-device-gen-fault |
General device fault |
event |
measurement |
1 |
8417909.11 |
inr-sensor-calibration-due |
The INR sensor calibration is due |
event |
measurement |
1 |
8408608.0 |
device-status-undetermined |
Undetermined device status event |
event |
measurement |
1 |
8408608.1 |
device-status-reset |
Device reset |
event |
measurement |
1 |
8408608.5 |
device-status-error |
Device error |
event |
measurement |
1 |
8408608.6 |
device-status-error-mechanical |
Mechanical error |
event |
measurement |
1 |
8408608.7 |
device-status-error-electronic |
Electronic error |
event |
measurement |
1 |
8408608.8 |
device-status-error-software |
Software error |
event |
measurement |
1 |
8408608.9 |
device-status-error-battery |
Battery error |
event |
measurement |
1 |
8408608.15 |
device-status-service |
Undetermined device status event, service needed |
event |
measurement |
1 |
8408608.16 |
device-status-service-time-sync-required |
Time sync service required |
event |
measurement |
1 |
8408608.17 |
device-status-service-calibration-required |
Calibration required |
event |
measurement |
1 |
8408608.18 |
device-status-service-replenishment-required |
Replenishment required |
event |
measurement |
1 |
8408608.25 |
device-status-battery-low |
Battery power is too low |
event |
measurement |
1 |
8408608.26 |
device-status-battery-depleted |
Battery is depleted |
event |
measurement |
1 |
8408608.27 |
device-status-battery-replaced |
Battery has been replaced |
event |
measurement |
1 |
8408608.28 |
device-status-battery-interrupted |
Battery was interrupted |
event |
measurement |
1 |
8418060.0 |
sensor-session-stopped |
Session stopped |
event |
measurement |
1 |
8418060.2 |
sensor-type-incorrect |
Sensor type incorrect for device |
event |
measurement |
1 |
8418060.3 |
sensor-malfunction |
The sensor malfunctioned |
event |
measurement |
1 |
8418060.4 |
device-specific-alert |
Device specific alert |
event |
measurement |
1 |
8418060.7 |
sensor-calibration-not-allowed |
Calibrating the sensor is not allowed |
event |
measurement |
1 |
8418060.8 |
sensor-calibration-recommended |
Calibrating the sensor is recommended |
event |
measurement |
1 |
8418060.9 |
sensor-calibration-required |
Calibrating the sensor is required |
event |
measurement |
1 |
8418060.10 |
sensor-temp-too-high |
Temperature is too high for a valid result |
event |
measurement |
1 |
8418060.11 |
sensor-temp-too-low |
Temperature is too low for a valid result |
event |
measurement |
1 |
8418060.12 |
sensor-result-below-patient-low |
Sensor result below the Patient Low level |
event |
measurement |
1 |
8418060.13 |
sensor-result-above-patient-high |
Sensor result above the Patient High level |
event |
measurement |
1 |
8418060.14 |
sensor-low-hypo |
Sensor result below level |
event |
measurement |
1 |
8418060.15 |
sensor-high-hyper |
Sensor result above level |
event |
measurement |
1 |
8418060.16 |
sensor-rate-decrease-exceeded |
The sensor rate of decrease has been exceeded |
event |
measurement |
1 |
8418060.17 |
sensor-rate-increase-exceeded |
The sensor rate of increase has been exceeded |
event |
measurement |
1 |
8418060.18 |
sensor-result-too-low |
Sensor result is lower than the device can process |
event |
measurement |
1 |
8418060.19 |
sensor-result-too-high |
Sensor result is higher than the device can process |
event |
measurement |
1 |
8418060.20 |
sensor-com-out-of-range |
The sensor communication is out of range |
event |
measurement |
1 |
8418512.0 |
Battery-status-Undetermined |
the status of the battery is unknown |
state |
measurement |
1 |
8418512.1 |
Battery-absent |
Indicates if the battery is absent |
state |
measurement |
1 |
8418512.2 |
Battery-active |
Indicates if the battery is active in the system |
state |
measurement |
1 |
8418512.3 |
Battery-charging |
Indicates if the battery is charging |
state |
measurement |
1 |
8418512.4 |
Battery-fullyCharged |
Indicates if the battery is fully charged |
state |
measurement |
1 |
8418512.5 |
Battery-disposable |
Indicates if the battery is disposable |
state |
measurement |
1 |
8418512.6 |
Battery-rechargeable |
Indicates if battery is rechargable |
state |
measurement |
1 |
8418512.8 |
Battery-faulty |
Indicates if battery is faulty |
event |
measurement |
2 |
8418512.7 |
Battery-overTemperature |
Indicates if the battery is too hot |
event |
measurement |
1 |
8418512.9 |
Battery-incompatible |
Indicates if battery is incompatible |
event |
measurement |
1 |
8410608.0 |
body-movement |
Indicates too much body movement |
event |
measurement |
1 |
8410608.1 |
cuff-too-loose |
Indicates the cuff was too loose |
event |
measurement |
1 |
8410608.2 |
irregular-pulse |
Indicates an irregular pulse was detected |
event |
measurement |
1 |
8410608.3 |
pulse-over-range-limit |
Indicates the pulse was too high for sensor to work |
event |
measurement |
1 |
8410608.4 |
pulse-under-range-limit |
Indicates the pulse was too low for sensor to work |
event |
measurement |
1 |
8410608.5 |
improper-body-position |
Indicates the cuff was incorrectly placed with respect to the heart |
event |
measurement |