Personal Health Device Implementation Guide
2.0.0-ballot - STU2 Ballot International flag

This page is part of the Personal Health Device FHIR IG (v2.0.0-ballot: STU 2 Ballot 1) based on FHIR (HL7® FHIR® Standard) v4.3.0. The current version which supersedes this version is 1.0.0. For a full list of available versions, see the Directory of published versions

CodeSystem: ASN1ToHL7

Official URL: http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 Version: 2.0.0-ballot
Active as of 2018-09-18 Computable Name: ASN1ToHL7

Maps IEEE 11073 ASN1 BITS measurements to codes. All these codes have two possible values, set or cleared. The code may either represent an event or a state. When an event, the defined event has occurred when set. If a state, the described item is either set or cleared. Sometimes the states may be defined in unexpected manners. For example, the regulation state when set is unregulated and when cleared it is regulated. In addition the code may derive from an IEEE 11073 attribute or from a measurement. More codes maybe added to this list in the future.

This Code system is referenced in the content logical definition of the following value sets:

Generated Narrative: CodeSystem ASN1ToHL7

Properties

This code system defines the following properties for its concepts

NameCodeURITypeDescription
Type of bit - indicates if a bit represents an Event (that happens) or a State (that persists) type http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7Concepts#type code Indicates that the code type is an event or state
Source - indicates if a bit comes from a measurement or from a device source http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7Concepts#source string Indicates that the code originates from a measurement or an attribute that further describes a measurement or the sensor device.

Concepts

This case-sensitive code system http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 defines the following codes in a Grouped By hierarchy:

LvlCodeDisplayDefinitionType of bit - indicates if a bit represents an Event (that happens) or a State (that persists)Source - indicates if a bit comes from a measurement or from a device
1 67925.0 onMains Indicates whether the PHD is on or off mains power state attribute
1 67925.1 onBattery Indicates whether the PHD is on or off Battery power state attribute
1 67925.8 chargingFull Indicates whether the PHD battery is charging at maximum rate event attribute
1 67925.9 chargingTrickle Indicates whether the PHD battery is charging at a trickle rate event attribute
1 67925.10 chargingOff Indicates whether the PHD battery charging is on or off state attribute
1 532354.0 regulation-status Indicates device is regulated when cleared state attribute
1 68219.0 mds-time-capab-real-time-clock Indicates real time clock is supported event attribute
1 68219.1 mds-time-capab-set-clock Indicates support for the PHG to set the time event attribute
1 68219.2 mds-time-capab-relative-time Indicates Relative Time is supported event attribute
1 68219.3 mds-time-capab-high-res-relative-time Indicates Hi-Res Relative Time is supported event attribute
1 68219.4 mds-time-capab-sync-abs-time Indicates that syncing the absolute time is supported event attribute
1 68219.5 mds-time-capab-sync-rel-time Indicates that syncing the relative time is supported event attribute
1 68219.6 mds-time-capab-sync-hi-res-relative-time Indicates that syncing hi-res relative time is supported event attribute
1 68219.7 mds-time-capab-bo-time Indicates that the real-time clock supports base offset time event attribute
1 68219.8 mds-time-state-abs-time-synced Indicates the absolute time has been synced event attribute
1 68219.9 mds-time-state-rel-time-synced Indicates the relative time has been synced event attribute
1 68219.10 mds-time-state-hi-res-relative-time-synced Indicates the hi-res time has been synced event attribute
1 68219.11 mds-time-mgr-set-time PHG Requested to set the time event attribute
1 68219.12 mds-time-capab-sync-bo-time Indicates that syncing base-offset time is supported event attribute
1 68219.13 mds-time-state-bo-time-synced Indicates that base-offset time has been synced event attribute
1 68219.14 mds-time-state-bo-time-UTC-aligned Indicates that base-offset time is aligned with UTC time standard event attribute
1 68219.15 mds-time-dst-rules-enabled Indicates support for daylight savings time rules event attribute
1 67846.0 lim-alert-off Indicates both the high and the low limit alerts are disabled state measurement
1 67846.1 lim-low-off Indicates the low limit alert is disabled state measurement
1 67846.2 lim-high-off Indicates the high limit alert is disabled state measurement
1 150604.0 sensor-disconnected Indicates the device is not connected with the sensor event measurement
1 150604.1 sensor-malfunction Indicates the sensor has malfunctioned event measurement
1 150604.2 sensor-displaced Indicates the sensor is not correctly placed on the user event measurement
1 150604.3 sensor-unsupported Indicates the sensor connected to the device is unsupported event measurement
1 150604.4 sensor-off Indicates the sensor is not connected to the user event measurement
1 150604.5 sensor-interference Indicates the sensor is experiencing interference event measurement
1 150604.6 signal-searching Indicates the sensor is searching for a signal event measurement
1 150604.7 signal-pulse-questionable Indicates a questionable pulse has been detected event measurement
1 150604.8 signal-non-pulsatile Indicates a non pulsatile signal has been detected event measurement
1 150604.9 signal-erratic Indicates a signal has been discovered, but it is erratic event measurement
1 150604.10 signal-low-perfusion Indicates the signal is experiencing low perfusion event measurement
1 150604.11 signal-poor Indicates the signal has been discovered, but it is poor event measurement
1 150604.12 signal-inadequate Indicates the signal has been discovered, but it is inadequate event measurement
1 150604.13 signal-processing-irregularity Indicates the signal has been discovered, but is experiencing processing irregularity event measurement
1 150604.14 device-equipment-malfunction Indicates a device malfunction has occurred event measurement
1 150604.15 device-extended-update Indicates an extended display update is in progress on device event measurement
1 150605.0 pulse-qual-nominal Indicates no abnormalities have been detected by the sensor event measurement
1 150605.1 pulse-qual-marginal Indicates the pulse profusion or quality is marginal event measurement
1 150605.2 pulse-qual-minimal Indicates the pulse profusion or quality is minimal event measurement
1 150605.3 pulse-qual-unacceptable Indicates the pulse profusion or quality is unacceptable event measurement
1 8410584.0 leadwire-loss Loss of unspecified lead wire or electrode connection event measurement
1 8410584.1 leadsignal-loss Loss of unspecified lead signal event measurement
1 8410584.2 leadwire-loss-first-lead Loss of 1st lead wire or electrode connection event measurement
1 8410584.3 leadsignal-loss-first-lead Loss of 1st lead signal event measurement
1 8410584.4 leadwire-loss-second-lead Loss of 2nd lead wire or electrode connection event measurement
1 8410584.5 leadsignal-loss-second-lead Loss of 2nd lead signal event measurement
1 8410584.6 leadwire-loss-third-lead Loss of 3rd lead wire or electrode connection event measurement
1 8410584.7 leadsignal-loss-third-lead Loss of 3rd lead signal event measurement
1 8417752.0 device-battery-low The device battery is low event measurement
1 8417752.1 sensor-malfunction The sensor has malfunctioned event measurement
1 8417752.2 sensor-sample-size-insufficient The sensor's sample size is insufficient event measurement
1 8417752.3 sensor-strip-insertion The test strip was inserted incorrectly event measurement
1 8417752.4 sensor-strip-type-incorrect The test strip type is incompatable with the sensor event measurement
1 8417752.5 sensor-result-too-high The value reported by the sensor is too high event measurement
1 8417752.6 sensor-result-too-low The value reported by the sensor is too low event measurement
1 8417752.7 sensor-temp-too-high The temperature is too high event measurement
1 8417752.8 sensor-temp-too-low The temperature is too low event measurement
1 8417752.9 sensor-read-interrupt The read process was interrupted event measurement
1 8417752.10 device-gen-fault General Device Fault event measurement
1 8417752.11 sensor-temp-out-of-range The temperature is out of range event measurement
1 8417909.0 inr-device-battery-low The INR device battery is low event measurement
1 8417909.1 inr-sensor-malfunction The INR device sensor has malfunctioned event measurement
1 8417909.2 inr-sensor-sample-size-insufficient The sample size was insufficient event measurement
1 8417909.3 inr-sensor-strip-insertion The test script was inserted incorrectly event measurement
1 8417909.4 inr-sensor-strip-type-incorrect The test strip used is not compatible with the INR event measurement
1 8417909.5 inr-sensor-result-too-high The test result is too high event measurement
1 8417909.6 inr-sensor-result-too-low The test result is too low event measurement
1 8417909.7 inr-sensor-temp-too-high The temperature is too high event measurement
1 8417909.8 inr-sensor-temp-too-low The temperature is too low event measurement
1 8417909.9 inr-sensor-read-interrupt The read process was interrupted event measurement
1 8417909.10 inr-device-gen-fault General device fault event measurement
1 8417909.11 inr-sensor-calibration-due The INR sensor calibration is due event measurement
1 8408608.0 device-status-undetermined Undetermined device status event event measurement
1 8408608.1 device-status-reset Device reset event measurement
1 8408608.5 device-status-error Device error event measurement
1 8408608.6 device-status-error-mechanical Mechanical error event measurement
1 8408608.7 device-status-error-electronic Electronic error event measurement
1 8408608.8 device-status-error-software Software error event measurement
1 8408608.9 device-status-error-battery Battery error event measurement
1 8408608.15 device-status-service Undetermined device status event, service needed event measurement
1 8408608.16 device-status-service-time-sync-required Time sync service required event measurement
1 8408608.17 device-status-service-calibration-required Calibration required event measurement
1 8408608.18 device-status-service-replenishment-required Replenishment required event measurement
1 8408608.25 device-status-battery-low Battery power is too low event measurement
1 8408608.26 device-status-battery-depleted Battery is depleted event measurement
1 8408608.27 device-status-battery-replaced Battery has been replaced event measurement
1 8408608.28 device-status-battery-interrupted Battery was interrupted event measurement
1 8418060.0 sensor-session-stopped Session stopped event measurement
1 8418060.2 sensor-type-incorrect Sensor type incorrect for device event measurement
1 8418060.3 sensor-malfunction The sensor malfunctioned event measurement
1 8418060.4 device-specific-alert Device specific alert event measurement
1 8418060.7 sensor-calibration-not-allowed Calibrating the sensor is not allowed event measurement
1 8418060.8 sensor-calibration-recommended Calibrating the sensor is recommended event measurement
1 8418060.9 sensor-calibration-required Calibrating the sensor is required event measurement
1 8418060.10 sensor-temp-too-high Temperature is too high for a valid result event measurement
1 8418060.11 sensor-temp-too-low Temperature is too low for a valid result event measurement
1 8418060.12 sensor-result-below-patient-low Sensor result below the Patient Low level event measurement
1 8418060.13 sensor-result-above-patient-high Sensor result above the Patient High level event measurement
1 8418060.14 sensor-low-hypo Sensor result below level event measurement
1 8418060.15 sensor-high-hyper Sensor result above level event measurement
1 8418060.16 sensor-rate-decrease-exceeded The sensor rate of decrease has been exceeded event measurement
1 8418060.17 sensor-rate-increase-exceeded The sensor rate of increase has been exceeded event measurement
1 8418060.18 sensor-result-too-low Sensor result is lower than the device can process event measurement
1 8418060.19 sensor-result-too-high Sensor result is higher than the device can process event measurement
1 8418060.20 sensor-com-out-of-range The sensor communication is out of range event measurement
1 8418512.0 Battery-status-Undetermined the status of the battery is unknown state measurement
1 8418512.1 Battery-absent Indicates if the battery is absent state measurement
1 8418512.2 Battery-active Indicates if the battery is active in the system state measurement
1 8418512.3 Battery-charging Indicates if the battery is charging state measurement
1 8418512.4 Battery-fullyCharged Indicates if the battery is fully charged state measurement
1 8418512.5 Battery-disposable Indicates if the battery is disposable state measurement
1 8418512.6 Battery-rechargeable Indicates if battery is rechargable state measurement
1 8418512.8 Battery-faulty Indicates if battery is faulty event measurement
2   8418512.7 Battery-overTemperature Indicates if the battery is too hot event measurement
1 8418512.9 Battery-incompatible Indicates if battery is incompatible event measurement
1 8410608.0 body-movement Indicates too much body movement event measurement
1 8410608.1 cuff-too-loose Indicates the cuff was too loose event measurement
1 8410608.2 irregular-pulse Indicates an irregular pulse was detected event measurement
1 8410608.3 pulse-over-range-limit Indicates the pulse was too high for sensor to work event measurement
1 8410608.4 pulse-under-range-limit Indicates the pulse was too low for sensor to work event measurement
1 8410608.5 improper-body-position Indicates the cuff was incorrectly placed with respect to the heart event measurement