Release 5

This page is part of the FHIR Specification (v5.0.0: R5 - STU). This is the current published version. For a full list of available versions, see the Directory of published versions . Page versions: R5 R4B R4

Example SpecimenDefinition/2364 (Narrative)

Orders and Observations Work GroupMaturity Level: N/AStandards Status: InformativeCompartments: No defined compartments

This is the narrative for the resource. See also the XML, JSON or Turtle format. This example conforms to the profile SpecimenDefinition.


Generated Narrative: SpecimenDefinition

Resource SpecimenDefinition "2364"

url: http://example.com/specdef/v1/12345

identifier: id: 12345

title: Example Specimen Definition for Testing

status: active

experimental: false

typeCollected: Venous blood specimen (SNOMED CT#122555007)

patientPreparation: 12 hour fasting (), At rest (SNOMED CT#263678003)

timeAspect: preferrably morning time

typeTested

isDerived: false

type: Serum specimen (SNOMED CT#119364003)

preference: preferred

Containers

-MaterialTypeCapMinimumVolume[x]
*plastic (SNOMED CT#61088005)Evacuated blood collection tube, thrombin/clot activator/gel separator (SNOMED CT#702281005)yellow cap (unknown#yellow)2 ml (Details: UCUM code mL = 'mL')

handling

temperatureQualifier: Ambient temperature ()

temperatureRange: 15-25 °C

MaxDurations

-ValueUnitSystemCode
*60minuteUnified Code for Units of Measure (UCUM)min

handling

temperatureQualifier: Refrigerated temperature ()

temperatureRange: 2-8 °C

MaxDurations

-ValueUnitSystemCode
*8hourUnified Code for Units of Measure (UCUM)h

typeTested

type: Plasma specimen (SNOMED CT#119361006)

preference: alternate

Containers

-MaterialTypeCapMinimumVolume[x]
*glass (SNOMED CT#32039001)Evacuated blood collection tube, thrombin/clot activator/gel separator (SNOMED CT#702281005)green cap (unknown#green)2 ml (Details: UCUM code mL = 'mL')

rejectionCriterion: insufficient specimen volume (RejectionCriterion#insufficient), hemolized specimen (RejectionCriterion#hemolized)

handling

temperatureQualifier: Ambient temperature ()

temperatureRange: 15-25 °C

MaxDurations

-ValueUnitSystemCode
*60minuteUnified Code for Units of Measure (UCUM)min

handling

temperatureQualifier: Refrigerated temperature ()

temperatureRange: 2-8 °C

MaxDurations

-ValueUnitSystemCode
*8hourUnified Code for Units of Measure (UCUM)h

 

 

Usage note: every effort has been made to ensure that the examples are correct and useful, but they are not a normative part of the specification.